start › cdms › k100 › run_diary › 20180829

2018/08/29 (Wednesday)

Previous Day

Setup:
DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer 3/?/2 → HVConn 1/3/2 → HV 1/2 → DCRC55(1)/DCRC50(3)

1.1kV → NaI B/C → Ortec 452(gain=5, shaping = 6 us, output=10V) / Ortec 575A(gain=10, shaping = 3us) → x10/x10 attenuator → DCRCX(4)
DCRC33(2) → kepco 1kV power supply → HV input
status: no data taken overnight.
0852 Start 1 x 2hr 40 V background datasets each split into 10min subruns, with CDMS and NaI detectors running and DCRC2 as HV controller. Using sequence HVscan-40×1-15_5PreBias-30Flash 30wait-2hrData-10minSubRun.msl
CDMS detector: Driver Gains = 2x, QET biases = (280/260/305)/(140/280/295) uA, Front-End Gains = 10
NaI detector: 1.1kV → NaI B/C → Ortec 452(gain=5, shaping = 6 us, output=10V, Pos) / Ortec 575A(gain=10, shaping = 3us, Pos) → x10/x10 attenuator → DCRC4PB/PA, Driver Gains = 1/1, PhononWaveFormLength = 1024 (512 samples), PhononPrepulseSamples = 64 (64 samples)
1000 BOR randoms, trigger on DCRC3pB/D threshold = 7/9
1025 MIDAS is taking data but nothing is being copied from /livedata to /byseries. Why? Stop sequence after finishing 9 10-minute series.
1030 Tower/MC/Coldplate = 30.3/29.3/- mK, Still = 1.3±0.1 K / 0.070 mbar, G1/G2 = 98/91 mbar, 1K Pot = 1.61 K / 20 mbar, Turbo on IVC = 1.0 nbar, MB = 36.9%
1030 run change_HV_40_to_0_DCRC2.msl
1035 Fill LN2 cold trap full.
1035 Flash DCRCs 1&3 for 30 seconds
1040 Kill then restart MIDAS. Turn on DAQ programs. On Messages can see that lazy logger is copying files.
1125 Reverse leads on HV power supply, we'll try -40V bias data.
1127 Start 1 x 2hr -40 V background datasets each split into 10min subruns, with CDMS and NaI detectors running and DCRC2 as HV controller. Using sequence HVscan-40×1-15_5PreBias-30Flash 30wait-2hrData-10minSubRun.msl
CDMS detector: Driver Gains = 2x, QET biases = (280/260/305)/(140/280/295) uA, Front-End Gains = 10
NaI detector: 1.1kV → NaI B/C → Ortec 452(gain=5, shaping = 6 us, output=10V, Pos) / Ortec 575A(gain=10, shaping = 3us, Pos) → x10/x10 attenuator → DCRC4PB/PA, Driver Gains = 1/1, PhononWaveFormLength = 1024 (512 samples), PhononPrepulseSamples = 64 (64 samples)
1000 BOR randoms, trigger on DCRC3pB/D threshold = 7/9
1140 Leakage looks a bit worse than in positive bias. Now Vbias = -40.013 V, Vtest = -6.77 mV. Typically in +bias Vtest is less than 5 mV.
1430 Start LHe transfer @ MB = 31.2%
1524 End LHe transfer @ MB = 93.7%
1605 Bring PuBe source back into lab and place centered behind lead stack at West wall of lead/poly shield (usual place)
The -40V data seemed to have a slightly lower noise width than the 40V data so we'll try taking some data at negative polarity for a while.
1610 Fill LN2 cold trap bath
1614 Start 7 x 2hr -40 V PuBe datasets each split into 10min subruns, with CDMS and NaI detectors running and DCRC2 as HV controller. Using sequence HVscan-40×1-15_5PreBias-30Flash 30wait-2hrData-10minSubRun.msl
CDMS detector: Driver Gains = 2x, QET biases = (280/260/305)/(140/280/295) uA, Front-End Gains = 10
NaI detector: 1.1kV → NaI B/C → Ortec 452(gain=5, shaping = 6 us, output=10V, Pos) / Ortec 575A(gain=10, shaping = 3us, Pos) → x10/x10 attenuator → DCRC4PB/PA, Driver Gains = 1/1, PhononWaveFormLength = 1024 (512 samples), PhononPrepulseSamples = 64 (64 samples)
1000 BOR randoms, trigger on DCRC3pB/D threshold = 7/9

Next Day