2018/08/22 (Wednesday)
| Setup: DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer 3/?/2 → HVConn 1/3/2 → HV 1/2 → DCRC55(1)/DCRC50(3) 1.1kV → NaI B/C → Ortec 452(gain=5, shaping = 6 us, output=10V) / Ortec 575A(gain=10, shaping = 3us) → DCRCX(4) DCRC33(2) → kepco 1kV power supply → HV input |
|
| status: Took SiHV+NaI PuBe data overnight. But for the entire sequence (starting at 07180821_1544), DCRC3pC&D were railed. | |
| 0845 | Tower/MC/Coldplate = 30.6/29.3/182 mK, Still = 1.4±0.1 K / 0.070 mbar, G1/G2 = 130/115 mbar, 1K Pot = 1.61 K / 20 mbar, Turbo on IVC = 1.0 nbar, MB = 71.5% |
| 0850 | Fill LN2 cold trap bath |
| 0855 | MIDAS webpage is down. Restart MIDAS. Sequence resumes and ramps bias up to 40V. Stop sequence after bias is ramped up. Adjust DCRC3pC/D driver offsets. Turn bias back down to 0V and flash detector. |
| 0957 | Start 1 x 2hr 40 V PuBe datasets each split into 10min subruns, with CDMS and NaI detectors running and DCRC2 as HV controller. Using sequence HVscan-40×1-15_5PreBias-30Flash 30wait-2hrData-10minSubRun.msl CDMS detector: Driver Gains = 2x, QET biases = (280/260/305)/(140/280/295) uA, Front-End Gains = 10 NaI detector: 1.1kV → NaI B/C → Ortec 452(gain=5, shaping = 6 us, output=10V, Pos) / Ortec 575A(gain=10, shaping = 3us, Pos) → x10/x10 attenuator → DCRC4PB/PA, Driver Gains = 1/1, PhononWaveFormLength = 1024 (512 samples), PhononPrepulseSamples = 64 (64 samples) 1000 BOR randoms, trigger on DCRC3pB/D threshold = 7/9 |
| 1052 | MB = 68.5% |
| 1122 | MB = 67.9% |
| 1149 | MB = 67.1% |
| 1226 | MB = 66.2% |
| 1258 | MB = 65.4% |
| 1320 | Start 7 x 2hr 40 V PuBe datasets each split into 10min subruns, with CDMS and NaI detectors running and DCRC2 as HV controller. Using sequence HVscan-40×7-15_5PreBias-30Flash 30wait-2hrData-10minSubRun.msl CDMS detector: Driver Gains = 2x, QET biases = (280/260/305)/(140/280/295) uA, Front-End Gains = 10 NaI detector: 1.1kV → NaI B/C → Ortec 452(gain=5, shaping = 6 us, output=10V, Pos) / Ortec 575A(gain=10, shaping = 3us, Pos) → x10/x10 attenuator → DCRC4PB/PA, Driver Gains = 1/1, PhononWaveFormLength = 1024 (512 samples), PhononPrepulseSamples = 64 (64 samples) 1000 BOR randoms, trigger on DCRC3pB/D threshold = 7/9 |
| 1650 | Fill LN2 cold trap bath |