2018/08/19 (Sunday)

Previous Day

Setup:
DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer 3/?/2 → HVConn 1/3/2 → HV 1/2 → DCRC55(1)/DCRC50(3)

1.1kV → NaI B/C → Ortec 452(gain=5, shaping = 6 us, output=10V) / Ortec 575A(gain=10, shaping = 3us) → DCRCX(4)
DCRC33(2) → kepco 1kV power supply → HV input
status: Took SiHV+NaI background data overnight.
1200 Tower/MC/Coldplate = 30.9/-/- mK, Still = 1.3±0.1 K / 0.075 mbar, G1/G2 = 133/127 mbar, 1K Pot = 1.62 K / 22 mbar, Turbo on IVC = 1.0 nbar, MB = 31.3%
didn't bother to record MC and Coldplate temperatures because a flash just happened.
1203 Stop data sequence after flash. I think we got 6x2hr series in.
1205 Fill LN2 cold trap bath.
1215 Commence LHe transfer, MB=31.3%
1303 Cease LHe transfer, MB=91.2%
Ending the transfer spiked the tower a little from 30.9 to 31.7 mK, settling around 31.4 mK now. Hopefully it comes back down.
1312 Circulation pressures have decreased slightly since transfer, now G1/G2 = 124/121 and falling slowly. 1K Pot wiggled at start and end of transfer, but is back to where it started now.
1315 Start 7 x 2hr 40 V background datasets each split into 10min subruns, with CDMS and NaI detectors running and DCRC2 as HV controller. Using sequence HVscan-40×7-15_5PreBias-30Flash 30wait-2hrData-10minSubRun.msl
CDMS detector: Driver Gains = 2x, QET biases = (280/260/305)/(140/280/295) uA, Front-End Gains = 10
NaI detector: 1.1kV → NaI B/C → Ortec 452(gain=5, shaping = 6 us, output=10V, Pos) / Ortec 575A(gain=10, shaping = 3us, Pos) → x10/x10 attenuator → DCRC4PB/PA, Driver Gains = 1/1, PhononWaveFormLength = 1024 (512 samples), PhononPrepulseSamples = 64 (64 samples)
1000 BOR randoms, trigger on DCRC3pB/D threshold = 7/9