2018/08/11 (Saturday)

Previous Day

Setup:
DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer 3/?/2 → HVConn 1/3/2 → HV 1/2 → DCRC55(1)/DCRC50(3)

1.1kV → NaI B/C → Ortec 452(gain=5, shaping = 6 us, output=10V) / Ortec 575A(gain=10, shaping = 3us) → DCRCX(4)
DCRC33(2) → kepco 1kV power supply → HV input
status: Took SiHV+NaI PuBe data overnight. Vbias very close to 40 V, Vtest around 5 mV.
Data from 07180810_1505 to 07180810_2137 looks good, but starting at 07180810_2147, DCRC1pB/3pC railed low. So the first 53% of the data is OK.
1000 Tower = 33.1 mK, Still = 0.090 mbar, G1/G2 = 119/111 mbar, 1K Pot = 1.64 K / 30 mbar
1140 Dial HV up to 40V with 15% prebias using appropriate script
1237 Start test data series to remotely adjust driver offsets
1324 Stop data. Adjusted offsets on DCRC1pB/3pC, -0.27/0.29 → -0.5/0.1. Now they aren't railed.
1329 Dial HV down to 0 using appropriate sequence.
1355 Flash detector (DCRCs 1,3) for 30 seconds
1430 Start 5 x 2hr 40 V PuBe datasets each split into 10min subruns, with CDMS and NaI detectors running and DCRC2 as HV controller. Using sequence HVscan-40×6-15_5PreBias-30Flash 30wait-2hrData-10minSubRun.msl
CDMS detector: Driver Gains = 2x, QET biases = (280/260/305)/(140/280/295) uA, Front-End Gains = 10
NaI detector: 1.1kV → NaI B/C → Ortec 452(gain=5, shaping = 6 us, output=10V, Pos) / Ortec 575A(gain=10, shaping = 3us, Pos) → DCRC4PB/PA, Driver Gains = 1/1, PhononWaveFormLength = 1024 (512 samples), PhononPrepulseSamples = 64 (64 samples)
1000 BOR randoms, trigger on DCRC3pB/D threshold = 7/9
1450 Fill LN2 cold trap bath. MB = 59.0%