status: Trying to stretch LHe bath over the weekend.
0912 MC(Cernox) = 5.7 K, Sorb/1K Pot = 9.4/>6.99 K, Turbo on IVC = 1.0 nbar, G1(fridge)/G2 = 559/509 mbar. P1/P2 = 1000/3.4 mbar. MB = 24.1%
0915 Fill LN2 cold trap full.
0925 Begin LHe transfer @ MB = 24.1%
0945 Start 4He pump. Pump out 1K Pot in steps, alternating between V4a and the 1K Pot speedivalve. Pot started at 0.37 bar, then fell each cycle to 0.18/0.10/0.06/0.03 bar. Pot temperature dipped each time to as low as 3.4 K, then settles at 3.9 K. Needle valve is still closed.
Sorb, MC, fridge pressure all fall.
1000 MC(Cernox)/Sorb/1K Pot = 5.6/4.8/3.9 K, G1/G2 = 443/404 mbar
1005 Open 1K Pot needle valve → 10°. 1K Pot → 3.1 K / 3.2 mbar
1010 Open 1K Pot needle valve → 20°. 1K Pot → 2.8 K / 5.0 mbar
Open 1K Pot needle valve → 30°. 1K Pot → 2.6 K / 10 mbar
MC starts falling very quickly, 5.3→4.1 K
1015 Open 1K Pot needle valve → 40°. 1K Pot → 2.3 K / 19 mbar
1020 MC(Cernox)/Sorb/1K Pot = 3.1/4.0(min?)/2.2 K, G1/G2 = 319/301 mbar. MC and fridge pressure falling.
Turbo on IVC = 0.74 nbar
1055 MC(Cernox)/Sorb/1K Pot = 2.1/4.0(min?)/1.5 K, G1/G2 = 58/57 mbar. MC and fridge pressure falling more slowly now.
Turbo on IVC = 0.86 nbar
1055 1K Pot = 1.47 K / 5.5 mbar.
Open 1K Pot needle valve → 50°. 1K Pot → 1.5 K / 7 mbar
Open 1K Pot needle valve → 60°. 1K Pot → 1.55 K / 11 mbar
1105 Spontaneous jump in 1K Pot → 1.7 K / 100 mbar
1150 MC(Cernox) = 1.89 K, G1/G2 = 49/42 mbar. 1K Pot = 1.69 K / 75 mbar
1157 Close V3 and V9, leaving 42 mbar in the dump.
1200 Start circulation through Trap 2. V6 → 12.8%. Condenser peaks at 94 mbar. Tower falls to 1.82 K. Small response in 1K Pot
1205 V6 → 13.2%. Condenser peaks at 115 mbar. Tower flat at 1.82 K (bad Lakeshore range) Tower < 1.82 K
1210 1K Pot falls to 1.66 K / 45 mbar
1213 V6 → 13.4%. Condenser peaks at 118 mbar. Tower flat at 1.82 K (bad Lakeshore range) Tower < 1.82 K
1216 V6 → 13.6%. Condenser peaks at 139 mbar. Increase Lakeshore resistance range, Tower = 1.33 K and falling. P1 falls within range. P1 = 12 mbar
1222 V6 → 14.0%. Condenser peaks at 120 mbar. Tower → 1.07 K, Still → 5.0 mbar
1226 V6 → 14.4%. Condenser flattens at 124 mbar. Switch bask to tower thermometer. Tower → 1.02 K, Still → 3.0 mbar
1230 V6 → 15.0%. Condenser flat at 126 mbar. Tower → 862 mK, Still → 1.6 mbar
1231 1K Pot has been flat at 1.66 K / 40 mbar
1236 V6 → 100%. Condenser flat at 129 mbar. Tower → 511 mK, Still → 0.24 mbar
1250 Tower → 85 mK, thermometer goes out of Lakeshore range again.
1308 Tower/MC/Coldplate = 40.4/33.5/182 mK, Still = 1.25±0.1 K / 0.085 mbar, G1/G2 = 78/81 mbar, 1K Pot = 1.66 K / 40±5 mbar. [image not archived: cool.svg]
1323 Purge E-box bag with a blast of N2 gas, then start slow flow of N2 gas.
1324 Tower = 36.9 mK and falling, Still = 0.095 mbar
1331 Open V14 to let the remaining mix out of the dump and into circulation. P1 peaks at 60 mbar, tower peaks at 87 mK
1341 Tower = 53 mK and cooling, Still = 3.0 mbar and falling
1345 Tower fell to 46.6 mK, then bounced up a little bit. P2 = 1.6 mbar
1356 P2 = 0.85 mbar, close V14
1415 Tower = 36.3 mK, Still = 0.110 mbar, G1/G2 = 77/79 mbar, 1K Pot = 1.67 K / 45 mbar
1430 Plug cables into HV box. Causes a temperature spike, Tower → 38.5 mK then cools again.
Plug in DCRCs. Start DAQ programs. Tower temperature spikes to 220 mK, then starts cooling again.
Setup: DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer 3/?/2 → HVConn 1/3/2 → HV 1/2 → DCRC55(1)/DCRC50(3)
1445 Tower = 37.8 mK and cooling.
1445 Zap SQUIDs on DCRCs 1&3. This seems to have failed: no temperature change, and at the end two of the DAQ programs switched themselves off.
DCRC6 was enabled by mistake, it's not plugged in. Change ODB then restart DAQ programs.
1455 Zap SQUIDs on DCRCs 1&3.
1505 Tower = 37.0 mK, can see pulses on all channels!
1508 Start 30 min mini-bake using flashandflash tool.
1555 Tower = 45.6 mK and falling
1600 Fill LN2 cold trap bath
1630 Start HV scan sequence of 60-minute background data series at HV = 0,20,40,60 V, Driver Gains = 2x, QET biases = (280/260/305)/(140/295/300) uA, Front-End Gains = 10, 1000 initial randoms, trigger on DCRC3pB/C/D threshold = 8/16/16, 30 s flashing then waiting 30 min after each series, 15% prebias for 5min, using sequencer script HVscan-0_20_40_60-15_5PreBias-30Flash-30wait-60data.msl
1632 1K pot = 1.66 K / 40 mbar
1800 At HV = 19.6 V, Vtest = 17 mV ⇒ Vdet = 12.8. Turn up N2 purge gas slightly, retape purge bag. test resistor voltage begins to fall slowly.
1810 Vtest = 13.3 mV and falling slowly.
1822 HV = 19.6 V, Vtest = 4.30 mV, Vdet = 17.9 V
2303 Start HV scan sequence of 60-minute background data series at HV = 0,20,40,60 V, Driver Gains = 2x, QET biases = (280/260/305)/(140/295/300) uA, Front-End Gains = 10, 1000 initial randoms, trigger on DCRC3pB/C/D threshold = 8/16/16, 30 s flashing then waiting 30 min after each series, 15% prebias for 5min, using sequencer script HVscan-0_20_40_60-15_5PreBias-30Flash-30wait-60data.msl