| 1315 | Perform mixture throughput test at MC = 77 K using above procedure. Let G1 rise to 590 mbar before Step 3, then it falls to 85 mbar, then begins to rise. Throughput 46.6 mbar/min. Close to April 25. |
| 1320 | Main bath = 0.4 mbar. Now let's fill it with He, see if the leak changes. Leak = 250 pbarL/s. |
| 1325 | Let about 0.5 bar He into IVC - leak signal rises above 10,000 pbarL/s. Stop letting in He; let's pump it out again and test some more. |
| 1325 | Leak = 18,000 pbarL/s. Start pumping He out of main bath again |
| 1340 | Main bath = 0.55 mbar, Leak = 331 pbarL/s. |
| 1340 | Let 1 hose-length of He into main bath. Leak → 355 pbarL/s. after a couple minutes, Leak = 375 pbarL/s |
| 1345 | Let a blast of He into main bath. Leak → 600 pbarL/s. after a couple minutes, Leak = 875 pbarL/s |
| | well this definitely looks like an IVC leak. But why didn't we see it when we blasted He down the fill tube?? |
| 1350 | Let a blast of He into main bath. Leak → 1100 pbarL/s. after a couple minutes, Leak = 875 pbarL/s |
| 1405 | Pump out main bath again, then fill with N2 again |
| 1425 | Main bath full of N2. IVC Leak = 225 pbarL/s. Try squirting blasts of He down the ports. Try the top port, then try the transfer tube at various heights, 15“ down to 1”, then try transfer tube pushed into funnel. For everything except tube-in-funnel, leak only increases by 1 pbarL/s at most. A blast down the tube with it pushed into the funnel causes a leak increase of 230→300 pbarL/s. A longer blast (count of 3) causes a 305→360 pbarL/s increase. These aren't huge. But it's clear that the leak is near the bottom of the bath. And the leak gets very large when you fill the bath with He. (Even when it's still well below 1atm, so it can't be a leak at the top from He leaking out of the bath). |
| 1615 | Try again with leak checks at top of fridge - no big leaks, maybe increases of +20 pbarL/s. Squirt down the fill line, extended blast; signal increase is maybe double this value. |
| 1620 | Pump out IVC again. |
| 1640 | Start letting He into IVC in bits. |
| | 1 hose-length of He, Main bath 7→9.5 mbar, IVC leak 200 pbarL/s, no increase |
| | 5 hose-length of He, Main bath →9.5 mbar, IVC leak → 400 pbarL/s |
| | more He, Main bath →2/30 bar, IVC leak → 2400 pbarL/s |
| | more He, Main bath →5/30 bar, IVC leak → 5000 pbarL/s |
| | more He, Main bath →15/30 bar, IVC leak → 15,000 pbarL/s |
| | more He, Main bath →338 mbar above 1 atm, IVC leak → 20,000 pbarL/s |
| | ok, just seems like too much. start warmup. |
| 1650 | MC(Cernox) = 77.5 K. Sorb = 142.8 K. |
| 1700 | Let N2 into the IVC up to about 0.21 bar. MC has climbed to 78.1 K |
| 1715 | Let 26 plastic-hose-lengths of N2 into cryostat vacuum jacket (equivalent of about 200 mbar at pump manifold). Vacuum jacket pressure = 10 mbar. Notice that main bath pressure increases. |
| 1715 | MC(Cernox) = 79.4 K, Sorb = 154.6 K |
| 1715 | Pump mixture to dump |
| 1745 | MC(Cernox) = 83.1 K, Sorb = 175.3 K |
| 1745 | Release main bath pressure to recovery |