| | Status: Tower assembled, ready to install. Mixture cleaned several times. Cryostat vacuum jacket pumped out. |
| 0925 | Check pressure in warm LN2 Trap 1: 6.4 mbar. Pump out with scroll pump. |
| 0930 | Check pressure in warm LN2 Trap 2: 4.5 mbar. Pump out with scroll pump. |
| 0940 | Start cryocleaning mixture through cold LN2 Trap 2, V6 = 12.6% |
| 0950 | Increase V6→12.9% to restart flow |
| 0950 | Increase V6→13.0% to speed up flow |
| 1015 | Increase V6→13.2% to speed up flow. Trap return tube gets frosty up past the o-ring flange. G2 maxxed at 700 mbar then falling slowly. |
| 1050 | Stop cryocleaning. Trap 2, 70 minutes. G2 = 645 mbar. Pump mixture back to dump. |
| 1235 | Start cryocleaning mixture through cold LN2 Trap 1, V6 = 13.2% |
| 1400 | Increase V6→13.4% to restart flow |
| 1450 | Stop cryocleaning. Trap 1, 135 minutes. G2 = 592 mbar. Pump mixture back to dump. |