2018/04/06 (Friday)
| Status: Warming fridge. All mixture in dump. 1K Pot needle valve open. Pumping out cryostat vacuum jacket overnight. Fridge and LN2 traps pumped out. IVC full of N2. | |
| 0000 | MC(Cernox) = 273.1 K. |
| 0400 | MC(Cernox) = 274.1 K. |
| 1005 | MC(Cernox) = 275.2 K. |
| Warming time constant climbed to 60 hours, 90 hours, 120 hours as vacuum jacket was pumped out. | |
| 1005 | Turbo pump = 8.6 nbar. Cryostat vacuum jacket = 160 nbar with pump valved out. Close vacuum jacket valve. |
| 1100 | Check pressures in LN2 Traps 1/2: 2.2/1.2 mbar. These were pumped out yesterday so this was a check of how much leaks into them overnight if they are isolated. Not much - not enough to account for the overnight rise in contaminants we saw during cryocleaning, with the dump open to (either) cold trap overnight. So if there's a leak in from the outside, it's probably in the connections between the dump and IGH or 3He pump and IGH. Yesterday I tightened all the accessible clamps at the dump, pump, and their IGH ports that had some give. |
| 1140 | Start cryocleaning mixture through cold LN2 Trap 1. V6 = 12.2% (strangely low). |
| 1145 | Increase V6→12.8% to restart flow |
| 1155 | Increase V6→12.9% to restart flow |
| 1210 | Increase V6→13.0% to accelerate flow |
| 1250 | Stop cryocleaning, 70 min through Trap 1. Pump mixture to dump. |
| 1245 | Close 1K Pot needle valve |
| 1255 | Lower main bath, erect clean tent |
| 1315 | Start cryocleaning mixture through cold LN2 Trap 2. V6 = 12.8% |
| 1320 | Increase V6→13.1% to restart flow |
| 1350 | Increase V6→13.2% to restart flow |
| 1425 | Stop cryocleaning, 70 min through Trap 2. Pump mixture to dump. |
| 1445 | Check pressure in warm LN2 Trap 1: 9.5 mbar. Seems to be stable with time (maybe falling a little) so trap is I think completely warm. Pump out with scroll pump. |
| 1500 | Deinstalling tower. At this point, just starting to remove 100mK stage |
| 1520 | 100mK stage removed. Ready to remove tower. |
| 1520 | Start cryocleaning mixture through cold LN2 Trap 1. V6 = 13.1% |
| 1525 | Increase V6→13.2% to restart flow |
| 1550 | Increase V6→13.3% to restart flow |
| 1600 | Modify SQUET ZC-24 with wirebonder. Now SQUIDs C and D have 6.7 mΩ shunts. This is meant to be a replacement for ZC-31 for use on DIB4 of the HV detectors. We had a suspicion that the ZC-31 SQUIDs were damaged during Run 66 (although bench checks show no problems). |
| 1615 | Perform electrical check on ZC-24, looks good |
| 1615 | Increase V6→13.4% to restart flow |
| 1630 | Stop cryocleaning, 70 min through Trap 1. Pump mixture to dump. |
| 1645 | Check pressure in warm LN2 Trap 2: 6.6 mbar. Pump out with scroll pump. |
| 1655 | Open dump to cold LN2 Trap 2 through V4 and V12b+speedivalve. Note this bypasses the pump which I'm leaving isolated. |