| | Setup: DIB1/3/4 → SL 27/19/20 → 3.2 MHz filters → cables 2,1,3 → HV board inputs 1,3,2 → HV board outputs 1,2 → DCRC55(1)/50(3). |
| 0701 | Begin IbIs data sequence on DCRC1pC with HV=0V from 750 to 400 uA in -0.5 uA steps, 3 seconds of data per step |
| | The sequence got stuck apparently - at 08:27 files stopped being produced, the last one at 526.5 uA. |
| 1020 | As far as I can tell (using the Opera browser through x-forwarding at home) the MIDAS website is unresponsive. I'll restart it again. |
| 1035 | Kill MIDAS, restart MIDAS and programs. As with the last hangup, had to go to ODB:RunInfo and set Transition In Progress to 0 because it was stuck in transition. Deleted the data files from /livedata from this partially-completed sequence, will restart it. |
| 1040 | Begin IbIs data sequence on DCRC1pC with HV=0V from 750 to 400 uA in -0.5 uA steps, 3 seconds of data per step |
| 1320 | Begin IbIs data sequence on DCRC3pB with HV=0V from 750 to 400 uA in -0.5 uA steps, 3 seconds of data per step |
| 1450 | Again, it looks like the sequence got stuck. Each time I've been able to run one IbIs sequence, but the second one gets stuck. This one stopped producing data at 14:42 at 526.5 uA - that's the same as the last time it got stuck! Something's up. |
| 1500 | Kill MIDAS, restart MIDAS and programs. Another detail with these hangups is that the kill_daq shell script does not kill mhttp, dcrcdriver, towerfe, triggerfe - these have to be killed individually with kill -KILL. |