| | Status: New oil in 3He pump. There is some air/N2 in the pump, and it is isolated from most of the IGH (V6 is open 100%). Pump is off. There is mixture in the Dump, Trap 2, the G1 volume, the IGH Manifold, the Vent volume, the Cryopump tubing, and the Cryopump. Cryopump is warm and open to G1, Trap 2, and Dump. |
| 0850 | G1 (dump+cryopump+Trap 2) = 882.0 mbar. |
| 0850 | P1 (3He pump inlet) = 0.4 mbar, G2 (3He pump outlet) = 119.5 mbar. 3He Pump oil level (pump off) = 60%. |
| 0850 | Turn on 3He pump. Initially rather loud, then gets quieter. No high-pitched whining. |
| 0855 | P1 (pump inlet) has fallen to 0.007 mbar. G2 (pump outlet) = 148.3 mbar. 3He Pump oil level (pump on) = 33%. |
| 0855 | Pump on 3He outlet through KF tee-port with scroll pump for about 15 seconds. New high-pitched noise from 3He pump, seems to be associated with low outlet pressure. I think this was true during the run, too. G2 (pump outlet) reduced to 77 mbar. 3He Pump oil level has changed a little, = 38%. |
| 0900 | Turn off 3He pump. Then continue pumping on 3He Pump outlet with scroll pump. G2 quickly falls to bottom of scale, 0.000 mbar. |
| 0905 | Open speedivalve to port at KF-tee to 3He pump inlet. Now we are pumping both sides with the scroll pump, and the pressure can be read on P1. P1 rises to 0.065 mbar, then begins falling. |
| 0910 | 3He pump oil level = 90%, with pump off and both sides being pumped on with the scroll pump. |
| 0910 | P1 (3He pump inlet and outlet) sorta flat at 0.060 mbar. Close off ports to 3He inlet and outlet, then turn 3He pump on again. P1 falls to 0.007 mbar. 3He pump oil level (pump on) = 60%, and bubbles seen in oil. |
| 0915 | G2 (3He pump outlet) rose to 2.2 mbar. Begin pumping on 3He pump outlet with scroll pump again. G2 falls to 0.000(minimum) again. |
| 0920 | Close speedivalve to stop pumping on 3He outlet. 3He pump still on, watch G2 to see if it rises above minimum. |
| 0925 | Can see a few bubbles in oil now but much less than before. |
| 0930 | G2 has not risen. Turn off 3He pump. Then with scroll pump still attached and on, open both speedivalves to connect 3He inlet and outlet so we can again see the pressure on P1. |
| 0930 | G2 briefly reads 0.2 mbar then falls back to minimum. P1 rises to 0.4 mbar, much higher than before. Are we getting out outgassing from the oil? P1 now falling as we continue to pump both sides of 3He pump out with scroll pump. |
| 0940 | P1 (3He Pump inlet+outlet) down to 0.030 mbar and continues to fall. When pumping it out a half hour ago it sorta fluctuated around 0.060, maybe because some of those air bubbles were slowly outgassing. |
| 1000 | P1 (3He Pump inlet+outlet) = 0.020 mbar and fairly flat. Stop pumping out 3He pump - close speedivalves to inlet and outlet and turn off scroll pump. |
| 1000 | Turn on 3He pump. Some bubbles seen in the oil but not nearly as many as at 9:10. |
| 1010 | Turn off 3He pump. Begin pumping outlet and inlet again with scroll pump. Forgot that pump was off before opening speedivalves: P1 rose to 10 mbar, G2 to 4 mbar. Both fall quickly when scroll pump turned on. |
| 1015 | P1 (3He Pump inlet+outlet) = 0.100 mbar and falling |
| 1140 | P1 (3He Pump inlet+outlet) = 0.010 mbar |
| 1145 | Stop pumping on 3He pump with scroll pump, close speedivalves at inlet and outlet. Turn on 3He pump. A few bubbles seen in oil. |
| 1155 | No more bubbles seen in oil. Turn off 3He pump. Start pumping inlet and outlet with scroll pump. P1 peaks at 0.200 mbar then falls. |
| 1305 | P1 (3He pump inlet+outlet) = 0.009 mbar. Stop pumping out 3He pump, close speedivalves to inlet and outlet. |
| 1305 | Turn on 3He pump. Bubbles seen in oil after first turning it on. |
| 1330 | No bubbles seen in 3He pump oil. Turn off 3He pump, then open the inlet&outlet speedivalves to start pumping on 3He pump with scroll again. P1 peaks at 0.270 mbar then falls. (No improvement over 11:55.) |
| 1350 | P1 (3He Pump inlet+outlet) = 0.013 mbar |
| 1350 | Stop pumping on 3He pump with scroll pump, close speedivalves at inlet and outlet. Turn on 3He pump. A few bubbles seen in oil. |
| 1420 | No bubbles. Turn off 3He pump again and open speedivalves to pump it out with scroll again. P1 peaks at 0.320 mbar then falls again. |
| 1425 | P1 down to 0.018 mbar. Stop pumping out with scroll then turn 3He pump on again. |
| 1420 | Turn off 3He pump again and open speedivalves to pump it out with scroll again. P1 peaks at 0.070 mbar then falls again. (Seems to be correlated with the amount of time the 3He is running.) |
| 1435 | P1 down to 0.016 mbar. Stop pumping out with scroll then turn 3He pump on again. |
| 1505 | Turn off 3He pump again and open speedivalves to pump it out with scroll again. P1 peaks at 0.260 mbar then falls again. (Better than the last 30 minute run at 14:20.) |
| 1520 | P1 down to 0.012 mbar. Stop pumping out with scroll then turn 3He pump on again. |
| 1550 | Turn off 3He pump again and open speedivalves to pump it out with scroll again. P1 peaks at 0.200 mbar then falls again. (Better than the last 30 minute run at 15:05.) |
| 1620 | P1 down to 0.010 mbar. Stop pumping out with scroll then turn 3He pump on again. |
| 1650 | Turn off 3He pump again and open speedivalves to pump it out with scroll again. P1 peaks at 0.140 mbar then falls again. (Better than the last 30 minute run at 15:50.) |
| 1710 | P1 down to 0.011 mbar. Stop pumping out with scroll pump, close speedivalves to 3He pump inlet & outlet. |