Status Fridge warming to room temperature. LN2 traps warm, one pumped out. Mixture in dump (LHe cold trap isolated at around 20 K). 1K Pot needle valve open. IVC filled with nitrogen.
0855 MC (Cernox) = 237.1 K
0855 Have been pumping out LN2 Trap 1 overnight, didn't intend to. Stop pumping out Trap 1.
0900 Open LN2 Trap 2 to G1: 27.3 mbar. Pump out Trap 2 with scroll pump.
1000 By Newton's law of cooling/warming, fridge won't reach dew point (283 K) until Wednesday. So I think I'll put some N2 gas in the cryostat vacuum jacket.
1010 Put one 3m hose length of N2 gas into cryostat vacuum jacket.
MC(Cernox) = 238.1 K
1035 No noticeable increase in warming rate. Put four more 3m hose lengths of N2 gas into cryostat vacuum jacket.
1055 Close 1K Pot needle valve
1200 OK, now fridge is warming more quickly. Previously 0.75 K/hr, for tau of about 78 hours - now 2.1 K/hr → tau = 27 hours. At this rate should hit dewpoint early on Sunday.
1210 MC (Cernox) = 241.1 K
1540 MC (Cernox) = 250.6 K, 3.0 K/hr, tau = 15 hours - faster now. Dewpoint could happen tomorrow morning.

S1502 bond pad photos

Wirebonding plan for S1502
1300 Carry S1502 to wirebonding lab
Do most of the above bonds. For DIB-to-detector use wirebonder settings: power 480/480, bond time 40/30, height 50, drop 26. For detector-to-detector: power 480/480, time 30/30, height 50, drop 10. Adjusted some of these upward because bonds weren't always working.
1530 Had problems with the wirebonder, the far end of the wire came off the spool, kind of a mess. Got to watch David from Dahlberg's lab rethread it completely, not just the needle. Will finish bonds later.