2017/08/13 (Sunday)

Previous Day

Status Fridge warming, circulation and 1K Pot off. Bath should be empty by now. LN2 traps empty and warm. Fridge open to dump. 1K Pot needle valve open. No longer pumping on IVC.
0645 MC (Cernox) = 21.20 K (checked remotely)
1005 MC (Cernox) / Sorb = 26.48 / 48 K (1K Pot maxed). IVC = 14 μbar. G1(Fridge&Dump) = 753 mbar and flat.
At 4:20-4:50 am Tower quickly rose from 4→20 K. Fridge pressure increased steadily until 4:30 am then flattened out.
1010 Start pumping fridge contents to dump
1020 G1 flat at 0.3 mbar, P1 flat at 0.016 mbar. G2(dump) = 774 mbar, still creeping up a bit.
Close Still gate valve and Condenser speedivalve to stop pumping out fridge. Close all GHS valves and turn off 3He pump.