2017/07/26 (Wednesday)

Previous Day

Setup: DIB 1/3/4 → 3.2 MHz filters 1/2/3 → cable 1/2/3 → gender changer 1/3/4 → DCRC55(1)/33(2)/X(4)
overnight: took 2/-2 V data triggering on a phonon channel. Last flash was at 04:42. Last data series began at 05:10. The sequence ends with grounding the detector but does not do a final flash, so the detector probably needs to be flashed again. Data looks good.
0900 Tower = 25.6 mK, Still = 0.060 mbar, G1/G2 = 163/152 mbar, 1K Pot = 1.69 K / 28 mbar. MB = 32.5%
0900 Lab = 23.4 C, Humidity = 43%
0905 Fill LN2 cold trap bath.
0925 Flash detector
1312 Tower/MC/Coldplate = 25.8/29.4/179 mK, Still = 1.1-1.5 K / 0.060 mbar, G1/G2 = 162/151 mbar, 1K Pot = 1.69 K / 30 mbar. MB = 25.9% (predicted level: 25.7%)
1323 Start LHe transfer (MB = 25.6%)
1404 End LHe transfer (MB = 94.2%)
1420 Release pressure from MB.
1622 Fill LN2 cold trap full.
1631 Start sequence 12 30-minute data series of charge-bias-scan data with Frontend Gain = (15/2)/(4/4)/(5/5/5/5), OG = 4/2/4, QET biases = 2V setting from 2017/07/25, with 1000 beginning randoms, 30 sec flash with 30 min wait, and a 15ADC bin threshold trigger on DCRC4pB (Side2pF). Charge biases are DCRC1/2 (Side2/1) = +/-. Side 2 biases in order are 0, 0.25, 0.5, 0.75, 1, 1.5, 2, 2.5, 3, 3.5, 4, 5