| 1000 | LD signal suddenly rises, to >2000 pbarL/s. Is this a leak near the bottom of the IVC? It seems like it happened pretty late, when the blowout was maybe already finished. In that case it could be He blowing into the lab and hitting the top of the stripline tube. Not sure. |
| 1005 | LD signal has reached 400 pbarL/s, leveling off |
| 1020 | Begin pumping out main bath with scroll pump. |
| | LD signal did not decrease when we started pumping out the MB (instead it's higher). If it was a leak in the bottom indium seal you may have expected the signal to fall at this point. |
| 1030 | Main bath (P2) is less than 1 mbar so it looks like all the LN2 is out. Stop pumping MB and backfill it with helium. |
| | It looks like maybe the leak signal did fall while pumping out the main bath - then while backfilling it increased a lot, to 29000 pbarL/s. So maybe there is a leak in one of the indium seals of the IVC. Another theory is that helium comes out at the top somewhere, maybe the stripline tube -to- main bath seal, then gets into the IVC through that tube. |
| 1045 | Let 1.2 mbar He exchange gas into IVC |
| 1050 | Let two long hose-lengths of He into cryostat vacuum jacket. |
| 1055 | At this point I've been pumping out the fridge lines for a while. Yesterday we did the throughput test through Trap 2. Today's test was through Trap 1. If there was somehow a lot of contamination in the fridge lines, it could show up in either/both of the traps now. We'll warm up Trap 2 to see, and for now do a cleaning through Trap 1. |
| 1100 | Remove Trap 2 from LN2 bath to let it warm up |
| 1110 | Start cryocleaning mixture through Trap 1. V6 at 13.2%. After a few minutes flow stops, increase V6 to 13.4% |
| | MC and Sorb are warming. Cryostat surface is cool to the touch. Taped the thermometer probe to the outside of the cryostat. |
| 1110 | MC(Cernox) = 78.8 K, Sorb = 135.4 K, Cryostat surface = 18.9 C, IVC = 1.4 mbar |
| 1130 | MC(Cernox) = 80.8 K, Sorb = 150.3 K, Cryostat surface = 17.2 C, IVC = 1.6 mbar |
| 1150 | Cryocleaning flow stopped, increase V6 to 13.6%. Later I noticed that this was probably too much, G2 rose over 900. Eventually it came down a bit. The line out of the cold trap got quite frosty. |
| 1200 | MC(Cernox) = 86.5 K, Sorb = 177.0 K, Cryostat surface = 16.3 C, IVC = 1.6 mbar |
| 1240 | Stop cryocleaning after 90 minutes. |
| 1245 | MC(Cernox) = 98.5 K, Sorb = 215.8 K, Cryostat surface = 16.5 C, IVC = 1.9 mbar |
| 1255 | Open Trap 2 (now close to room temperature) up to G1. About 7.6 mbar of contaminants, not very much. |
| 1345 | MC(Cernox) = 116.2 K, Sorb > 249.9 K (max), Cryostat surface = 16.8 C, IVC = 2.2 mbar |
| 1425 | MC(Cernox) = 129.2 K, Cryostat surface = 11.1 C, IVC = 2.5 mbar |
| 1440 | Try a throughput test @ 130 K, using Trap 2, same procedure as yesterday except never opened V9, so using a smaller amount of mixture, whatever was left in the pump. G1 reaches maximum of 220 mbar before isolating mix in still, then falls to minimum of 12 mbar after opening G1 to condenser. Again, no throughput. Throughput < 0.1 mbar/min. |
| 1507 | Open Trap 1 (now close to room temperature) up to G1. About 2.3 mbar of contaminants, not very much. |
| 1620 | MC(Cernox) = 161.6 K, Cryostat surface = 18.2 C, IVC = 3.0 mbar |
| 1640 | Using Newton's Law of Cooling (or warming) the current trend suggests MC will hit freezing (273 K) at 8:05 in the morning. |