2017/03/29 (Wednesday)

Previous Day

Setup: DIB1/3/4 → 3.2 MHz filters 3/1/2 → cable 3/1/2 →gender changer 3/1/2 → HVConn 1/3/2 → HV 1/2 → DCRC55/DCRC33
Fridge failure, the pump power is off. Looks like it happened some time before midnight, earlier than the LabView look-back time
0935 Tower 1.8 K, 1K Pot 5.6 K / “1 bar” (max), G1/G2 = 355/315 mbar, Still = “1 bar” (max)
0940 Fill LN2 cold trap bath
0940 Close 1K Pot pumping valve and v4a, so that when we turn the 4He pump back on it won't be overly taxed. Turn off both pumps on IGH control.
Reset pump power box in pump alley. Then turn on 4He pump.
Open V4a to pump out pot pumping line.
Open 1KPot speedivalve. Pot temperature falls quickly.
0945 Check MC temperature: MC = 2.0 K
1K Pot temperature and G1/G2 (condenser) pressures are falling
0950 1K Pot = 1.8 K / 28 mbar, G1/G2 = 220/195 mbar
Turn on 3He pump. Condenser pressure goes high, try to close v6 but too slow, G1 goes over 1 bar so I think some mix got diverted into the dump.
0955 With v6 closed, condenser pressure is settling, G1/G2 = 540/490 mbar and falling.
1030 1K Pot = 1.65 K / 16 mbar, G1/G2 = 80/65 mbar.
Open v6 to 13%. G1/G2 go up about 5 mbar.
Open v6 to 13.5%. G1/G2 rise above 125 mbar
1040 1K Pot = 1.69 K / 26 mbar, G1/G2 = 175/155 mbar and flat.
Open v6 to 14.2%. G1/G2 rise to 295/265 mbar
Still pressure begins to fall
1045 1K Pot = 1.70 K / 32 mbar, G1/G2 = 290/260 mbar. Condenser pressure and Tower temperature falling a bit
Still = 4 mbar. Open v6 to 15%
1050 MC = 1.3 K and falling
G1/G2 flattish at 295/265 mbar. Open v6 to 15.8%. Still → 0.55 mbar
1055 Main bath = 35.3%
1055 Tower/MC = 1.1/1.17 K, Still = 0.6 mbar, G1/G2 = 295/265 mbar and flat
Open v6 100%
1115 Tower/MC = 0.86/0.96 K, Still = 0.17 mbar, G1/G2 330/295 mbar and flat
Open v14 to pull any extra mix out of dump. Let G1/G2 rise to about 600 mbar then close v14 again. Surge in Still and 1K Pot
1130 Tower/MC = 275/230 mK, Still = 0.13 mbar, G1/G2 = 355/320 mbar and slowly falling
Open v14 again. Again close it after G1/G2 hit 600 mbar (takes longer this time).
1145 Tower/MC = 87/78 mK, Still = 0.12 mbar, G1/G2 = 325/295 mbar and slowly falling
Open v14 again. G1 peaks at 510. Still rises then falls to 16 mbar, will leave open until Still less than 1 mbar.
1210 Tower/MC = 47/44 mK, Still = 0.9 mbar, G1/G2 = 285/255 mbar and slowly falling
Close v14.
1245 Tower/MC/Coldplate = 33.7/31.3/191 mK, Still = 1.3 K / 0.120 mbar, G1/G2 = 255/230 mbar, 1K Pot = 1.706 K / 32 mbar, Main bath = 32.6%. Back to normal! Boil off rate: close to slow mode (predicted level: 30.2%).
1430 Commence LHe transfer, MB=29.5%
1530 Cease LHe transfer, MB=89.0%
1734 Fill LN2 coldtrap bath.