| | Setup: DIB1/3/4 → 3.2 MHz filters → cables → nothing/33(2)/DCRC55(1) |
| 1415 | Tower/MC/Coldplate = 44.8/32.2/190.4 mK, Still = 1.2 K / 0.110 mbar, G1/G2 = 146/141 mbar, 1K Pot = 1.67 K / 22 mbar, Main bath = 33.1%. Boil off rate: close to slow mode. |
| 1415 | Fill LN2 cold trap. |
| 1426 | 1K Pot (P2) = 24 mbar. Open 1K needle valve a few degrees. |
| 1441 | 1K Pot = 1.72 K / 38 mbar. |
| 1445 | Start LHe transfer. MB = 33.5% |
| 1525 | Finish LHe transfer, MB = 64.8%. Dewar had about 22L left. |
| 1540 | Release pressure from MB. |
| 1625 | Re release pressure from MB. |