start › cdms › k100 › run_diary › 20170126

2017/01/26 (Thursday)

Previous Day

Setup: DIB1/3/4 → 3.2 MHz filters → cables → DCRC55(1)/33(2)/50(3)
0840 Fill LN2 cold trap.
0845 Tower/MC/Coldplate = 62.0/33.2/192 mK, Still = 1.2 K / 0.110 mbar, G1/G2 = 131/129 mbar, 1K Pot = 1.70 K / 32 mbar, Main bath = 71.3%.
1150 Measure rate by counting pulses in 10*100ms traces on DIB1PA: 70Hz
1200 Try to measure rate in channel below source (DIB4PB) but find the channel does not respond to changes in QET bias.
Channel Roll Call
DIB Channel Responds to
QET bias changes
Responds to
Test signal on FB
Notes
DIB1PA Side1 pE yes yes
DIB1PB Side1 pB yes yes
DIB1PC Side1 pD yes yes
DIB1PD Side2 pE yes yes
DIB3PA Side2 pB yes no* Unresponsive to QET bias on Friday Jan 27
DIB3PB Side2 pA yes no*
DIB3PC Side1 pA yes
DIB3PD n/c slightly
DIB4PA Side2 pC no no
DIB4PB Side2 pF no no Responsive to QET bias using DCRC1 on Friday Jan 27. But not on Monday Jan 30.
DIB4PC Side1 pC yes Only weak pulses seen using DCRC1 on Friday Jan 27. Pulses look okay as of Feb 8.
DIB4PD Side1 pF yes
*This may not be functional on DCRC33

This seems to indicate that a couple of channels that should be connected, DIB4 pA and pB, are maybe open, because there is no change in signal with a change in QET bias. We tried looking at the test signal in feedback directly as a check of whether the DCRC seemed to be working for these channels, but we became a bit confused as to what we should see.

1245 Try IbIs at low biases on DCRC1pA. Up to 50 uA it looks like a standard s/c slope: calculated Rpar = 10.4 mΩ.
1300 Try applying test signals to DCRC1 QETs, 12 mVpp/100 Hz. At first DCRC1pA looked reasonable: output 22 mVpp, with driver gain 1x so this is around 10 mΩ. But DCRC1pC and pD show some pulses as if they are in transition… and looking later at DCRC1pA it's output is much lower, 8 mVpp → 70 mΩ, so it seems to have been driven normal.
1310 MB = 64.9 %. Boil off rate: faster than slow mode but not much.
1525 Try a different method for Tc. Set all DCRC1 QET biases to 10 μA. All seem to be superconducting. Driver gains = 1x. By watching the baseline you should see a change when the TES goes normal, the baseline should rise significantly (if Rpar = 10 mΩ and Rn+Rpar = 70 mΩ then baseline should rise by more than 100 mV). Between s/c and normal there should probably be a transition region in which pulses are seen.
1530 Tower = 60.9 mK.
1600 This method seemed to work. Applying 5-8 μW to the MC heater, scanned through a tower temperature range of 61-73 mK; all the transitions temperatures were in this range. Scanned up and down, could see the move into transition (pulses appear, baseline rises) and into normal (pulses mostly disappear, baseline reaches a maximum). Scanned up and down and recorded the data below. All Tc's on the DIB1 channels are between 68 and 71 mK.
S101 Critical Temperatures
DIB chan Side chan SC → Transition [mK] Transition → Normal [mK] Normal → Transition [mK] Transition → SC [mK]
DIB1 pA Side1 pE 65.8 68.7 69.3 69.3
DIB1 pB Side1 pA 67 70.4 69.4 70.2
DIB1 pC Side 1 pD 70 72.1 72.2 71.7
DIB1 pD Side 2 pE 69.3 71.2 71.1 71
1630 Fill LN2 cold trap bath
1635 Tower = 60.8 mK
1635 MB = 59.9%