2017/01/17 (Tuesday)

Previous Day

Status: Main bath raised, IVC turbo pumping. Small leak at top of stripline tube. Main bath vacuum jacket is evacuated. 1K Pot needle valve closed. LN2 traps clean. Mixture is in dump, fridge evacuated. Room temperature.
1100 Turbo pump = 8.6 nbar, IVC = 0.46 μbar
Preparing to transfer LN2
1130 Begin leak detector on IVC. Starts around 50 pbarL/s but it's falling. Falls to 2.0 pbarL/s in a couple of minutes.
1135 Pump out main bath with scroll pump. (Before pumping out bath we double checked that the 1K Pot was empty - it wasn't entirely. Pumped out about 3 mbar from pot.) Pumps down to below 1 mbar on P2 in less than 5 minutes.
1145 Pressurize MB with He gas. Leak detector signal rises to 15 pbarL/s.
1155 Let 4.6 mbar He exchange gas into IVC
1205 Begin transferring LN2 to main bath
MC(Cernox) = 300 K, Sorb = >249 K (max), IVC = 4.6 mbar.
1220 MC(Cernox) beginning to fall, 297 K
1305 MC(Cernox) = 240 K, Sorb = >249 K (max), IVC = 4.0 mbar. Level resistor = 1.05 kΩ
1305 LN2 storage Dewar is empty, need to refill it. Pause transfer.
1410 MC(Cernox) = 205 K, Sorb = >249 K (max), IVC = 3.0 mbar. Level resistor = 1.05 kΩ
1410 Re-start LN2 transfer
1420 MC(Cernox) = 200 K, Sorb = >249 K (max), IVC = 3.0 mbar. Level resistor = 1.12 kΩ
1455 MC(Cernox) = 170 K, Sorb = >249 K (max), IVC = 2.2 mbar. Level resistor = 1.20 kΩ. Getting frosty at top of fridge, plume of condensation coming out of the port.
1505 MC(Cernox) = 160 K, Sorb = 247 K, IVC = 1.9 mbar. Level resistor = 1.23 kΩ.
1520 MC(Cernox) = 140 K, Sorb = 221 K, IVC = 1.9 mbar. Level resistor = 1.28 kΩ.
1530 MC(Cernox) = 130 K, Sorb = 197 K, IVC = 1.6 mbar. Level resistor = 1.31 kΩ.
1540 MC(Cernox) = 150 K, Sorb = 178 K, IVC = 1.6 mbar. Level resistor = 1.326 kΩ. Close to finished.
1550 MC(Cernox) = 116 K, Sorb = 167 K, IVC = 1.6 mbar. Level resistor = 1.330 kΩ. Close to finished.
1555 MC(Cernox) = 112 K, Sorb = 157 K, IVC = 1.6 mbar. Level resistor = 1.330 kΩ.
Stop LN2 transfer
1630 MC(Cernox) = 96 K, Sorb = 136 K, IVC = 1.4 mbar.
1702 MC(Cernox) = 90 K, Sorb = 131 K, IVC = 1.4 mbar. Time to go so commence pumping out IVC with scroll pump.
1705 MC(Cernox) = 89 K, Sorb = 131 K, IVC = 6.3e-2 mbar. Switch to pumping on IVC with turbopump.

Next Day