| 1237 | Setup: DIB1/3/4 → 3.2 MHz filters → cables1/2/3 → DCRC55(1)/33(2)/50(3) |
| | Tower/MC/Coldplate = 33.4/35.9/232 mK, Still = 1.3 K / 110 μbar, G1/G2 = 197/172 mbar, 1K Pot = 1.7 K / 30 mbar. MB = 28.7%. |
| | Fill LN2 cold trap full. |
| 1246 | Commence LHe transfer (27.7%) |
| 1310 | Pause LHe transfer at 56.3% as dewar ran dry (7L remaining). Resume with another Praxair dewar. |
| 1319 | Remove 88Y source from safe and place in Pb castle. |
| 1323 | Enable DCRC4 for coincident data. |
| 1227 | Start test series with Frontend Gain = 5/2/5, OG = 8/3/8, with 1000 beginning randoms, trigger on PhA of DCRC4 (NaI trigger) at 10 ADC. |
| 1335 | DCRC1 PB was railed, adjust driver offset. |
| 1345 | Cease LHe transfer (89.8%) |
| 1353 | Start sequence of 64 1-hour series of 0V coincident data with Frontend Gain = 5/2/5, OG = 8/3/8, with 1000 beginning randoms, no flashing, trigger on PhA of DCRC4 (NaI trigger) at 10 ADC. |
| 1354 | Release pressure form main bath. |
| 1430 | Release pressure form main bath. |