Baked 12 hours overnight. Tower temp oscillated between 500mK and 600mK until bake ended at 0516 and then fell to base by ~0530. 1K pot has held steady at ~30mbar all night.
1346 Tower/MC/Coldplate = 37.5/36.6/196 mK, Still = 1.05 K / 85 μbar, G1/G2 = 184/183 mbar, 1K Pot = 1.70 K / 30.0 mbar. MB = 63.1%
Boiloff rate seems to be halfway between fast- and slow-mode
1359 Fill LN2 cold trap full.
1400 Open 1K pot needle valve open another 1 degree. 1K pot pressure settles at 45mbar.