| | DCRC1 and DCRC3 powered on and connected through HV setup EBoxDIB 2/3 → 3.2 MHz filters → Cables → remapping board I/remapping board II → HVside2board “DIB1”/“DIB2” → DCRC55(1)/50(3). Remapping boards have all possible lines switched off. . |
| | Baking using IR LEDs for 14 hours last night. Base temp held at 545mK until bake ended at 0350, then fell to ~90mK |
| 0900 | Tower/MC/Coldplate = 87.8/56/??(3.2 kΩ) mK, Still = 1.08 K / 95 μbar, G1/G2 = 196/188 mbar, 1K Pot = 1.57 K / 6.0 mbar. MB = 42.9% 1K Pot has been stable. |
| 0902 | Fill LN2 cold trap bath. |
| 1050 | After fiddling with the breadboard and resistor, we concluded that having the resistor on the breadboard attached with wires to the HV switchboard produced the lowest levels of white noise, so we will take data in that configuration. |
| 1110 | Check pulses and attempt to reduce trigger threshold as much as possible. 300 ADC Bins seems to be the lowest we can go without excess noise triggers. |
| 1130 | Flash with IR LEDs for 30s and cooldown for 10 minutes. |
| 1150 | With 33kΩ resistor on DIB1 QOFB line, manually start series of 30 minutes data with 20V HV/ -10V DIB1 with 1000 starting/1000 ending randoms with 300 ADC Bin trigger threshold on DIB1 Qo. |
| 1222 | Cease data run. Ground detector. Flash with IR LEDs for 30s, and cooldown for 10 minutes. |
| 1238 | With 33kΩ resistor on DIB1 QOFB line, manually start series of 30 minutes data with 20V HV/ -10V DIB1 with 1000 starting/1000 ending randoms with 300 ADC Bin trigger threshold on DIB1 Qo. |
| 1310 | Cease data run. Ground detector. Flash with IR LEDs for 30s, and cooldown for 10 minutes. |
| 1327 | With 33kΩ resistor on DIB1 QOFB line, manually start series of 30 minutes data with 40V HV/ -10V DIB1 with 1000 starting/1000 ending randoms with 300 ADC Bin trigger threshold on DIB1 Qo. |
| 1402 | Cease data run. Ground detector. Flash with IR LEDs for 30s, and cooldown for 10 minutes. |
| 1417 | With 33kΩ resistor on DIB1 QOFB line, manually start series of 45 minutes data with 40V HV/ -10V DIB1 with 1000 starting/1000 ending randoms with 300 ADC Bin trigger threshold on DIB1 Qo. |
| 1506 | Cease data run. Ground detector. Flash with IR LEDs for 30s, and cooldown for 10 minutes. |
| 1520 | With 33kΩ resistor on DIB1 QOFB line, manually start series of 45 minutes data with 60V HV/ -10V DIB1 with 1000 starting/1000 ending randoms with 300 ADC Bin trigger threshold on DIB1 Qo. |
| 1617 | Cease data run. Commence LHe transfer (31.9%). |
| 1637 | With 33kΩ resistor on DIB1 QOFB line, manually start series of coincidence test data with 12V HV/ -12V DIB1 with 1000 starting/1000 ending randoms with 1 ADC Bin trigger threshold on DIB4 PA which is connected to an external HPGe trigger. |
| 1642 | Cease data run. Ground detector. Flash with IR LEDs for 30s, and cooldown for 10 minutes. |
| 1659 | End LHe transfer (80.7%). |
| 1704 | Release pressure from main bath. |
| 1706 | Accidentally log some garbage data. |
| 1710 | With 33kΩ resistor on DIB1 QOFB line, manually start series of 60 minutes data with 60V HV/ -10V DIB1 with 1000 starting/1000 ending randoms with 300 ADC Bin trigger threshold on DIB1 Qo. Set to trigger on DIB1 Qinner, not Qouter, data is garbage |
| 1732 | Release pressure from main bath. |
| 1752 | Release pressure from main bath. |
| 1813 | Cease data run. Ground detector. Flash with IR LEDs for 30s, and cooldown for 10 minutes. |
| 1830 | With 33kΩ resistor on DIB1 QOFB line, manually start series of 60 minutes data with 60V HV/ -10V DIB1 with 1000 starting/1000 ending randoms with 300 ADC Bin trigger threshold on DIB1 Qo. Data has 1000 bin prepulse instead of 500 bin prepulse |
| 1933 | Cease data run. Ground detector. Flash with IR LEDs for 30s, and cooldown for 10 minutes. |
| 2002 | With 33kΩ resistor on DIB1 QOFB line, manually start series of 90 minutes data with 80V HV/ -10V DIB1 with 1000 starting/1000 ending randoms with 300 ADC Bin trigger threshold on DIB1 Qo. Data still bad. Unsure why |
| 2032 | Something is still wrong with this data. Stop series after a few moments, template length is still not matching up, though everything we normally check looks fine. Load ODB from previous series, before DAQ was altered. Check everything again and it looks ok, except it loaded as not triggering on anything, maybe because it ends on randoms? Set to trigger on DIB1 Qo. Ground detector. Flash with IR LEDs for 30s, and cooldown for 10 minutes. |
| 2059 | With 33kΩ resistor on DIB1 QOFB line, manually start series of 90 minutes data with 80V HV/ -10V DIB1 with 1000 starting/1000 ending randoms with 300 ADC Bin trigger threshold on DIB1 Qo. |
| 2128 | Something is still wrong with this data. Stop series after a few moments, template length is still not matching up, though everything we normally check looks fine. AHA, DCRC4 is enabled and powered. Disable and power down DCRC4. Flash with IR LEDs for 30s, and cooldown for 10 minutes. |
| 2239 | Sooo…. not sure why, but this was fixed after doing the following things. 1) Open squidGUI and start it to look at pulses. Then kill it. 2) Run flashandtake.py with -10V on DIB1 for 5 minutes. Data looked ok. Flash with IR LEDs for 30s, and cooldown for 10 minutes. |
| 2240 | Y-88 source put out. |
| 2250 | With 33kΩ resistor on DIB1 QOFB line, manually start series of 30 minutes data with 80V HV/ -10V DIB1 with 1000 starting/1000 ending randoms with 300 ADC Bin trigger threshold on DIB1 Qo. |
| 2322 | Cease data run. Ground detector. Flash with IR LEDs for 30s, and cooldown for 10 minutes. |
| 2334 | With 33kΩ resistor on DIB1 QOFB line, manually start series of 30 minutes data with 100V HV/ -10V DIB1 with 1000 starting/1000 ending randoms with 300 ADC Bin trigger threshold on DIB1 Qo. |
| 2339 | Y-88 source put away. |
| 1156 | Cease data run. Ground detector. Commence 8 hour IR bake. Fill LN2 cold trap full. |