| | Setup: iZIP mode, DIB1/2/3 → cables → 3.2 MHz filters → DCRC55/33/50 |
| | Taking background data overnight ended at 13:42 |
| 1547 | Fill LN2 cold trap Dewar |
| 1553 | Tower/MC(H35)/MC sensor/Cold Plate = 31.3/26.5/41/163 mK, Still = 1.23 K / 0.090 mbar, G1/G2 = 132/134 mbar, 1K Pot = 1.59 K / 4.0 mbar, Main bath = 30.3% |
| 1603 | Tweak 1K Pot needle valve open additional 5 degrees. |
| 1615 | Anthony Started 16 hours of 30 min bg series at -4/+4 V with standard trace lengths with FG=5/2/5, OG=3, with 1000 initial randoms, 1000 ending randoms, with 30s flash and 25 minute cooldown in between series. Trigger on DIB1 phA with threshold of 30ADC bins, with 150 us pileup cut window. |
| 1619 | Commence LHe transfer (29.5%) |
| 1721 | Finish LHe transfer (90.6%) |
| 1729 | Released main bath pressure for 30sec |