| | Setup: iZIP mode, DIB1/2/3 → cables → 3.2 MHz filters → DCRC55/33/50 |
| | Taking Y-Be photoneutron data overnight. Last flash at 07:26 |
| 0905 | Remove Y source from source holder and put back into safe. |
| 0910 | Tower/MC(H35)/MC sensor/Cold Plate = 31.1/26.9/41/162 mK, Still = 1.32 K / 0.11 mbar, G1/G2 = 160/135 mbar, 1K Pot = 1.68 K / 10 mbar, Main bath = 70.1% |
| | Tweak 1K Pot needle valve open a few more degrees |
| 0915 | Fill LN2 cold trap Dewar. |
| 1207 | Start 1 30min background series at -4/+4V with FG=5/2/5, OG=3, with 1000 initial randoms, 1000 ending randoms, 30s flash and 25 minute cooldown before series. Trigger on DIB1 phA with threshold of 20ADC bins and DIB1 qInner&qOuter each with threshold 210 ADC bins, with 150 us pileup cut window. |
| 1330 | Remove Be discs from source holder. Waiting until cooldown is almost done before putting out strong 88-Y source. |
| 1333 | Start 1 30min Strong 88-Y series at -4/+4V with FG=5/2/5, OG=3, with 1000 initial randoms, 1000 ending randoms, 30s flash and 25 minute cooldown before series. Trigger on DIB1 phA with threshold of 20ADC bins and DIB1 qInner&qOuter each with threshold 210 ADC bins, with 150 us pileup cut window. |
| 1357 | Place strong Y source in holder without Be (right at end of cooldown just before DAQ starts taking data) |
| 1445 | Remove Y source from fridge, return to safe. |
| 1445 | Put Cf source in shield. Planning to check raw event rate versus Cf position |