start › cdms › k100 › run_diary › 20160115

2016/01/15 (Friday)

Previous Day

Setup: iZIP mode, DIB1/2/3 → cables → 3.2 MHz filters → DCRC55/33/50
Taking Y-88 data overnight. Last flash at 07:34
0900 Tower/MC(H35)/MC sensor/Cold Plate = 31.9/26.3/41.2/165 mK, Still = 1.15 K / 0.12 mbar, G1/G2 = 145/150 mbar, 1K Pot = 1.71 K / 20 mbar, Main bath = 52.4%
0940 Fill LN2 cold trap Dewar.
1205 Place Cf source inside shield near detector for activation while performing DCRC maintenance.
1245 For DCRC55, the Qi intercept/Qi slope/Qo intercept/Qo slope calibration register values were set to 8067/01AB/8018/01AB. See instruction at MIDAS.
1505 Commence LHe transfer (39.2%).
1600 Finish LHe transfer (89.1%).
1607 For DCRC50, the Qi intercept/Qi slope/Qo intercept/Qo slope calibration register values were set to 80A3/01A8/8027/01AC. See instruction at MIDAS.
1620 Measured the bath level through the fill port with a thumper (borrowed from the LHe room). At main bath = 89.2%, the level is 58.9 cm below the top of the fill port.
this seems to put it near the top of the upper part of the IVC. Note: fill port top is approx. 2.75“ above main bath ceiling
1655 Place Cf source back in safe so that we take only 88-Y data over the weekend.
1700 After applying firmware calibration settings to DCRCs, attempt to look at pulses. Attaching DCRCs to Ebox causes temperature to spike. Along with zapping SQUIDs, temperature spikes up to ~68mK. Temperature falls quickly, but stabilizes around 50mK. Attempt switching back to not using calibrations, but this doesn't have an affect.
Attempt unpowering DCRCs one at a time. First DCRC55.
No effect.
Unplug DCRC33.
No effect.
Unplug DCRC50.
Effect.
1715 Fill LN2 cold trap full.
Power DCRC55. Small temperature spike from base, but only a few mK. Plug in DCRC50, slightly larger temperature spike, but settles back down. Plug in DCRC33. Everything seems fine. Zap SQUIDs. Here goes everything.
1740 Finish looking at channels and adjusting ADC Offsets. DIB2 pC is upside down. Tried relocking. No effect. Zap SQUIDs to see if that helps.
1800 Zapping fixed it, but then I had to readjust the ADC Offsets.
1808 Start 36 hours of 30min 88-Y series at -4/+4V with FG=5/2/5, OG=3, with 1000 initial randoms, 1000 ending randoms, 30s flash and 25 minute cooldown before series. Trigger on DIB1 phA with threshold of 30ADC bins.

Next Day