start › cdms › k100 › run_diary › 20151230

2015/12/30 (Wednesday)

Previous Day

Status: First LHe transfer was yesterday. MC (Cernox) cooling overnight, turbo pump pumping out IVC. Cf source in shield for activation purposes.
Both traps are at room temperature being pumped out by 4He pump. P2 = 0.018 mbar.
0930 Sorb/MC(Cernox)/Tower = 16.6/18.2/18.5 K. IVC pumping/valved out = 3.0e-7/1.6e-6 mbar. Main bath = 46.4%.
0940 Isolate IVC.
0945 Close GHS valves and 4He pump valve, turn off 4He pump
0950 Place cold traps in LN2 cold trap Dewar and fill LN2 cold trap Dewar.
0955 Let mix into Trap 1 from dump. Leaving 3He pump off. With V9 open and pump off, G2=758.3 mbar. Open V13a: G2 falls to 712 mbar.
1005 Let mix into fridge through Condenser valve, V1, and Trap 1 speedivalve. Also open gate valve to still to view pressure on P1. (This of course opens an additional large volume for mixture to expand into: the circulation pumping line.)
1010 G1/G2 = 610/635 mbar and falling. Sorb = 10.8 K and falling. MC(Cernox) = 17.8 K
1015 Open 1KPot needle valve 45 degrees. Sorb and 1KPot temperatures start to fall.
1020 Pot fell to 6.6 K but then rose again. Sorb fell to 8.0 K but also rose again. Sorb = 8.7 K. Still/Coldplate = 16.0/NaN(?)
cold plate resistance looks like a normal value but the vi doesn't translate it to a temperature, don't know why
1025 Close V13a to isolate some mix in the dump. G2 = 553 mbar.
1030 Start pumping on 1KPot. G1 dropping < 500 mbar, Sorb ~ 4K, 1KPot → 3.1 K / 6.0 mbar.
1045 Pot slowly cooling, G1 falling slowly. 1KPot = 2.15 K / 6.0 mbar. Still/MC(Cernox)/Tower = 5.6/17.5/17.8 K. Cold plate thermometer still not reading right.
1050 Open 1KPot needle valve another 5 degrees to increase cooling power (pumping pressure is still lowish, 6 mbar)
1110 1KPot = 1.72 K / 18.0 mbar. G1 = 228 mbar. MC(Cernox) = 16.4 K.
1115 Letting more mix into fridge in controlled fashion. Open V13A but watch 1KPot pressure, close again if it climbs high.
1120 Leave V13a open. G2 falling.
1125 1K Pot = 2.3 K / 18.0 mbar. G1/G2 = 380/440 mbar, getting closer to one another. Still/ColdPlate/MC(Cernox)/Tower = 3.7/6.1/14.1/14.2 K
1135 Open V12a completely, allow mix to flow freely into fridge. Small increases in G1 and 1KPot.
1145 G1/G2 = 405/365 mbar. Still/ColdPlate/MC(Cernox)/Tower = 3.2/4.9/3.24/4.51 K. 1K Pot = 2.2 K / 18.0 mbar.
1200 G1/G2 = 215/205 mbar. Still/ColdPlate/MC(Cernox)/Tower = 2.4/3.6/2.70/3.92 K. 1K Pot = 2.2 K / 18.0 mbar.
1210 Close V13a and turn on 3He circulation pump. This pushes mix that's in the pump into the dump. G2 rises from 160 to 168 mbar. Then reopen V13a.
1220 G1/G2 = 105/100 mbar. Still/ColdPlate/MC(Cernox)/Tower = 2.0/2.9/2.36/3.19 K. 1K Pot = 1.62 K / 7.0 mbar. Open needle valve additional 20 degrees.
1225 Close V9 leaving 80 mbar in dump.
1230 1KPot spikes for no apparent reason. Delayed response? Pressure settling around 75 mbar. Close needle valve about 10 degrees
1235 G1/G2 = 80/70 mbar. Still/ColdPlate/MC(Cernox)/Tower = 2.1/3.0/2.29/2.32 K. 1K Pot = 1.76 K / 34.0 mbar.
1240 Start circulation: open V6 to 12%. Then start opening further in steps of 0.5%.
1248 At V6 = 13.5% response is seen: Circulation and 1KPot pressures increase. Still temperature starts to decrease.
1250 G1/G2 = 225/250 mbar. Still/MC(Cernox) = 1.9/1.59 K. 1K Pot = 1.76 K / 34.0 mbar.
1255 At V6 = 14.0%, still pressure(P1) falls into readable range. Falls as low as 10 mbar, some oscillation.
1300 V6 = 14.0%, G1/G2 = 125/155 mbar. Still = 1.26 K / 20 mbar, ColdPlate/MC(H35)/Tower = 1.30/1.33/1.21 K. 1K Pot = 1.75 K / 30.0 mbar.
1300 Still pressure oscillations stop. Was spiking between about 5 and 50 mbar. Now staying below 5 mbar.
1315 V6 = 14.5%, G1/G2 = 125/120 mbar. Still = 1.09 K / 2.8 mbar, MC(H35) = 0.74 K.
1340 V6 = 16.5%, G1/G2 = 100/95 mbar. Still = 1.26 K / 0.380 mbar, MC(H35) = 0.153 K.
1340 V6 = 100%, G1/G2 = 100/95 mbar. Still = 1.09 K / 0.140 mbar, MC(H35) = 59 mK.
1350 Begin opening then closing V14 to let the rest of the mix into circulation. Throttle to keep circulation pressures between 80 and 160 mbar. MC has fallen as low as 30 mK but increases as high as 100 mK when more mix is let in.
1430 Leave V14 open to get all mix in fridge
1550 G1/G2 = 100/95 mbar. Still = 1.11 K / 0.240 mbar, V14 open. ColdPlate/MC(H35)/Tower = 151/28.9/40.3 mK.
1555 Almost all mix is out of dump, so close V14
1600 Commence LHe transfer (30.3%).
1645 G1/G2 = 100/95 mbar. Still = 1.11 K / 0.120 mbar, ColdPlate/MC(H35)/Tower = 150/28.3/39.5 mK. 1K Pot = 1.73 K / 28.0 mbar.
1710 Cease LHe transfer (92.8%).
1715 Fill LN2 cold trap full.

Next Day