Setup: iZIP mode, DIB1/2/3 → cables → 3.2 MHz filters → DCRC55/33/50
Taking Cf data overnight with MIDAS. Last flash at 0648. Charge biases now at zero - flash-and-take now does this at the end of the sequence.
0745 Small temperature spike to 33.5mK. Returned to base temperature after several minutes.
0900 Main bath = 6.0%. G1/G2: 109.6/107.5. P1: 100ubar, P2: 5.5mbar, 1KPot temperature: 1.595K, Tower temperature: 32.1mK. Close needle valve of 1K pot. Pump mixture to dump. P1 rises to 0.28mbar, then begins falling.
Fill LN2 cold trap full.
0905 1K Pot temperature drops to 1.404K, P2: 0.85mbar. Close 1K Pot speedivalve. Temperature rises to 1.6K.
0910 Tower temperature: 241mK. Turn on mixing chamber heater to 20mW. Turn the yellow valve on the He4 Pump exhaust from Recovery to Exhaust.
0915 P1 rises to 9.0mbar before dropping quickly back below 1mbar. Open V12A and Cold Trap 1 speedivalve to pump to dump whatever remaining mixture may still be in there.
0917 G2 shows 335mbar of mixture collected so far. Tower temperature at 1.38K, Mixing Chamber temperature at 1.489K.
0927 Turn off He4 Pump.
0937 Tower temperature at 1.29K. G2 shows 459mbar of mixture collected in dump.
1130 Tower temperature at 11.7K. G2 shows 799mbar of mixture collected in dump.
1255 Mixing Chamber temperature at 9.8K. Sorb temperature at 5.4K. Isolate Turbopump to measure pressure in line leading up to IVC: 8.6E-6mbar. Open IVC. Pressure at 1.2E-5mbar. Open turbopump to IVC and pressure begins dropping.
1300 G2 shows 798mbar of mixture collected in dump. G1 is below 1mbar. P1: 16ubar. Close V1 and V6 with Mixing Chamber heater still on.
1305 P1 doesn't rise from 16ubar. All mixture must be recovered from fridge. Open V1 and V6. Close Condensor Line Speedivalve and Gate Valve. Then close V1 and V6 again.
Main Bath = -0.1%. Close He Recovery valves, but leave Main Bath isolated from atmosphere.
Close V12A, V12B, and both Cold Trap Speedivalves. Remove cold traps from LN2 bath and open V11A, V11B, V7, V2A, and V5A to connect Cold Traps to P2.
1310 V6 closed all the way. Close V9 to isolate dump. Close manual valves on dump to further isolate mixture over break.
Turn off He3 pump.
1400 Traps are warm, and P2 reads 26mbar. Turn on He4 pump and commence pumping out traps.
1405 MC temperature = 7.5K.