Setup: iZIP mode, DIB1/2/3 → cables → 3.2 MHz filters → DCRC55/33/50
Taking background data overnight with MIDAS. Last flash at 07:35
0835 Tower/MC(H35)/MC sensor/Cold Plate = 32.0/27.4/42.4/170 mK, Still = 1.33 K / 0.11 mbar, G1/G2 = 115/135 mbar, 1K Pot = 1.61 K / 6.0 mbar, Main bath = 21.1%
0845 Start LHe transfer at 20.1%
0855 Fill LN2 cold trap Dewar
0855 Set charge biases back to 0/0 V.
1003 Finish LHe transfer at 88.0%
1115 Prepare to switch cold traps mid-circulation. G1/G2: 116.9mbar/136.1mbar, P1: 10ubar, Tower temperature: 32.4mK.
1116 Open V13B to allow mixture in Cold Trap 2.
1117 Open V12B and Cold Trap 2 speedivalve to bring Cold Trap 2 into circulation. Close V13A to stop circulating through Cold Trap 1, but continue pumping mixture out of it.
1120 Close V12A and Cold Trap 1 speedivalve. P1: 9ubar. Open V11A and V8 to pump out the rest of the mixture in Cold Trap 1 to the still side. P1 rises to 55mbar, starts dropping.
1150 P1: 10ubar. Close V8 and V11A. Circulation back to normal. Tower temperature had spiked when reintroducing mixture, but then it dropped below normal, and now it's risen slightly to 33mK. Watching it to see how it behaves.
1712 Place Cf source at fridge for overnight data taking.
1720 Saw some strong ~200Hz noise. Unplug bung from LHe dewar to release pressure. At least temporarily, noise goes away.
1723 Start 15 30min series of -/+4V MIDAS Cf data; FG=5/2/5, OG=3; with 30s flash and 25min cooldown in between series. Trigger on DIB1pA only (Side1pB) with trigger threshold = 30 and Pileup Cut threshold to 100 μs.
1725 Fill LN2 cold trap full.