UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20151208

2015/12/08 (Tuesday)

Previous Day

Setup: iZIP mode, DIB1/2/3 → cables → 3.2 MHz filters → DCRC55/33/50
Cf source near detector for overnight activation.
0815 Tower/MC(H35)/MC sensor/Cold Plate = 35.2/28.6/41.2/156 mK, Still = 1.08 K / 0.095 mbar, G1/G2 = 120/125 mbar, 1K Pot = 1.55 K / 4 mbar, Main bath = 24.0%
0830 Fill LN2 cold trap Dewar
0832 Start LHe transfer at 20.3%
0954 End LHe transfer at 90.4%
1120 1KPot is still running quite low. Try closing needle valve completely and reopening a couple of times. Open it to about 45 degrees.
1240 1K Pot temperature: 1.473K, P2: 2.2mbar. Still running low. I now realize that we haven't been recovering the exhaust of the He4 pump, which shouldn't be affecting why the 1K Pot pressure keeps dropping, but who knows. Switch He4 pump exhaust to recovery and see if that affects the pot.
1250 1K Pot temperature: 1.453K, P2: 1.5mbar. Looks like it's still dropping. Open needle valve another 1/8th of a turn.
Temperature spikes up to ~2K, and P2 goes off scale. Then drops back down to ~1.83K and 120mbar. Then starts rising. Looks like it's oscillating for now.
1300 1K Pot temperature has settled ~1.75K and P2 at ~33mbar. The tower temperature was gradually increasing while the 1K Pot dried up, but it dropped when the needle valve was opened, and it's held steady at ~36mK now.
1320 Place Cf source back in safe.
1325 1K Pot temperature has dropped to ~1.62K and P2 is at 6.0mbar. Tower temperature has risen to 38mK and is still gradually rising.
1407 1K Pot temperature continued to drop to ~1.58K, P2: 4.0mbar. Tower temperature had actually decreased to ~37.6mK. Open needle valve more to see if we can get to a point where the 1K pot stays higher.
1435 Start looking at pulses to see what state Midas left them in. On DIB 1, there's 250Hz noise. Let some pressure out of He dewar to see if that helps.
Initially it seems like the sinusoidal noise went away, but it comes back periodically.
1443 When switching to view pulses on DIB2 (DCRC33), Detector Readout wouldn't update the displays. Had to WR 15 0, WR 15 20 to telnet session in order to get displays to update. All four channels on DIB2 are railed.
1449 When switching to view pulses on DIB3 (DCRC50), Detector Readout wouldn't update the displays. Had to WR 15 0, WR 15 20 to telnet session in order to get displays to update. All phonon channels except phC on DIB3 are railed.
1452 250Hz noise still present on all DIBs. (LHe level at 86.4%)
1507 At some point during our 250Hz noise investigations, Detector Readout VI froze as it had done yesterday. It was looking at DIB3 (DCRC50) at the time. Abort and rerun the VI.
1530 Listening to the vibrations on the top of the fridge platform and comparing to a tone generator, estimate that the frequency there is about 104-105 Hz. Measured the frequency of the noise in the phonon pulses: it's close to 208 Hz, i.e. the first harmonic of the fridge vibration.
1535 Release pressure from top of main bath again, with level at 84.1%. Then plug it again. 250 Hz noise goes away, at least for a couple of minutes…
1538 Fridge frame no longer vibrating to the touch, and sinusoidal noise hasn't shown back up on pulses.
1540 Establish timing link w/ DIB1 (DCRC55) as master.
1542 1K Pot temperature had dropped to ~1.5K and P2 at 2.2mbar. Open needle valve even more.
1545 Set phonon trigger thresholds to following ADC bin values:
A B C D
DIB 1 32 30 40 38
DIB 2 18 18 22 22
DIB 3 56 42 40 60
1552 While setting DIB3 (DCRC50) phonon thresholds, Detector Readout VI freezes again.
1606 Sinusoidal oscillations still gone. LHe level at 83.8%.
1608 Start 8 cycles of background data alternating between series of -/+4V and +/-4V for a total of 16 series with 1000 initial randoms, 2000 events/file; 30s flash and 25min cooldown in between series.
1616 Stop data because event rate was bogged down. Attempted to raise thresholds by 5ADC bins each during the data run, but it didn't work well enough. Adjusted phonon trigger thresholds are below:
A B C D
DIB 1 38 36 46 40
DIB 2 18 24 24 22
DIB 3 56 50 40 60
1623 Start 8 cycles of background data alternating between series of -/+4V and +/-4V for a total of 16 series with 1000 initial randoms, 2000 events/file; 30s flash and 25min cooldown in between series.
1636 Rate was still too low, so after playing with phonon thresholds, set all phonon trigger thresholds to 64ADC bins and event rate jumps up to the expected 15Hz. So I guess there's still some tweaking to be done. Stop data.
A B C D
DIB 1 40 40 56 48
DIB 2 24 24 34 28
DIB 3 64 50 46 66
1645 Start 8 cycles of background data alternating between series of -/+4V and +/-4V for a total of 16 series with 1000 initial randoms, 2000 events/file; 30s flash and 25min cooldown in between series.
1655 It seems the event rate is averaging ~11Hz. Oscillating between 13 and 8Hz. Good enough to leave overnight.
1720 The 1K Pot has dropped back to ~2mbar, so open the needle valve even more during cooldown.
1725 Fill LN2 cold trap full.

Next Day