2015/09/26 (Saturday)
| Current setup: iZIP mode, DIB1/2/3 → cables → 3.2 MHz filters → DCRC55/33/50. | |
| Overnight: Took bkg data. Last flash was some time before 04:40. Tower temperature stable since then. | |
| 1635 | Fill LN cold trap Dewar |
| 1640 | Begin LHe transfer at 25.5% |
| 1700 | Tower/MC(H35)/MC sensor/Cold Plate = 30.8/25.3/41.0/158 mK, Still = 1.34 K / 0.085 mbar, G1/G2 = 155 mbar, 1K Pot = 1.72 K / 22 mbar |
| 1733 | Finish LHe transfer at 80.3% |