| | Current setup: iZIP mode, DIB1/2/3 → cables → 3.2 MHz filters → DCRC55/33/50. |
| | Overnight: Cf activation. Tower temperature stable. |
| 0855 | Fill LN cold trap Dewar |
| 0902 | Tower/MC(H35)/MC sensor/Cold Plate = 30.9/25.3/41.3/159 mK, Still = 1.38 K / 0.080 mbar, G1/G2 = 150 mbar, 1K Pot = 1.72 K / 24 mbar, Main bath = 34.8% |
| 0905 | Remove Cf source from fridge |
| 0915 | ~270 Hz sinusoidal noise seen in all phonons. Intermittent, usually not present but every few 10s of seconds it reappears to a maximum amplitude of maybe 3× the white noise (~20 mVpp) |
| 1040 | Modify LabVIEW DAQ to only wait on copying data for up to 5 minutes. |
| 1103 | Set up MultiDAQ to take 100 randoms at 0V with 6.5536ms trace length to test alternative trace lengths for processing. Start data. |
| | Rate low (~5Hz) and it also looks like it's triggering? So there's that… |
| 1106 | Set MultiDAQ to take 100 randoms at 0V with 50ms trace length. |
| 1108 | Start Data. Rate at ~1Hz and it seems like timing link might be down? |
| 1135 | Reset timing link just in case. |
| 1201 | No obvious noise frequencies. |
| 1204 | Take 20ms traces of properly biased QETs. 0V Qbias. 1000 randoms. Rate ~1.5Hz. |
| 1218 | Stop data. Set all QET biases to 0. Set to trigger on Qo=3 ADC bins in addition to phonon channels in hopes that the weird triggering we think we see in the randoms doesn't happen. |
| 1231 | Start 1000 randoms of superconducting traces of 20ms. ~1.1Hz event rate. Rate drops to ~0.2Hz. Disable triggering on Qo, rate bounces back to ~1.2Hz. |
| 1254 | Stop data. Set QET biases to 400uA across the boards. |
| 1305 | Start 1000 randoms of normal resistance data. Rate= ~1.2Hz. |
| 1334 | Stuck triggering, lower thresholds to 25 on DIB1. Change DIB1 pA threshold to 16ADC bins & then it begins triggering again. Rate= ~1.3Hz. |
| 1348 | Stop data. |
| 1411 | Revert to transition QET biases. Stack DCRC boards and take 1000 randoms of 20ms traces. |
| 1415 | Start Data. Rate= ~1.2Hz. |
| 1439 | Stop data. Unpower boards to remove filters. |
| 1458 | Repower boards. Zap SQUIDs using MIDAS. |
| 1511 | Zapping with MIDAS appears to work fine, except that it applies a -/+4V Qbias to the crystal. We'll need to address that. |
| | Have to adjust Driver Offsets because of filters: |