| 1415 | Move 10 MHz filter to DIB2. Now 10 MHz C-filter on DIB2 at ebox, HV1/2 → DCRC50/55. Note this means that 4 out of 8 channels on Side 2, the bias side, are now filtered. |
| 1420 | Leave Side1 charge bias at zero, increase HV in + direction. Above +15 V started losing phonons, but they would come back… slowly increased HV in ~0.5 V steps, which would cause HV1pA to rise 50 mV, then wait for it to relax again. Slowly raise HV… |
| 1445 | HV up to +26.3 V (49.6 nA), definitely higher than before… |
| 1505 | Reach Kepco 0-40V supply's maximum of +31.1 V (60.7 nA). Phonon pulses good, no change in tower temperature. Meter reads +31.03 V. Hence Vcrystal = +6.75 V. |
| 1510 | Apply opposite bias to Side1 charge channels, increase in -1 V steps. Reach -14 V with no problems. |
| 1520 | Try - HV now. First set all biases to 0. Then increase HV to -31 V (-60.4 nA) quickly - no problem at all with phonon pulses. |
| 1525 | Now increase Side1 charge bias to +14 V - phonon pulses disappear but begin to come back after a couple of minutes, but still small. |
| 1530 | HV1pA flat at its high-end baseline, HV1pB pulses very distorted, HV1pC seems OK. (This one behaves as if it has the highest Tc.) Return Side1 charge bias back to 0V, but pA still flatline, pB and pC pulse-less. HV had to be reduced to -28 V to make pulses return. |
| 1540 | Behavior different now, increasing HV seems to elevate temperature, raise baselines, kill pulses. Raising Side1 charge bias also raises baselines. At some voltages pulses return over time, but limiting voltages seem to be worsening. |
| 1615 | HV at -21.93/-22.22/-23.22 V (by multimeter), -39.1/-39.6/-41.5 nA, causes HV1pA baseline to increase to 0.275/0.3/0.34(max) V. |
| 1645 | Try the 0-1000 V Kepco DC supply. 60 Hz spikes in charge channels with ground side of HV plugged in to supply, 60 Hz sinusoid with bias side plugged in, if supply plugged directly into wall, or through Variac, or with groundless plug. But supply→Variac→groundless plug→wall works fine. |
| 1654 | HV = 35.00/45.41/50.00 V (by multimeter) , 69.8/93.5/103.8nA ⇒ V_crystal = 7.08/8.01/8.84V. Phonon pulses present. |
| 1705 | Fill LN2 cold trap full. |
| 1705 | Main bath = 47.1%. This is actually a very slow boiloff - only 1.1%/hour. |
| 1735 | HV = 55.48V, 116.2nA, V_crystal = 9V. Phonon pulses present. |
| 1742 | Cranked HV to 60V quickly. Lost phonons, charge pulses still good. |
| 1756 | Phonons beginning to return. HV to 70.00V (multimeter) causes MC temp to go from 31.6mK to 32.2mK, but seems stable. Charge pulses still okay. |
| 1802 | HV to 80.00V (multimeter), 169.9nA. MC temp increase to 32.6mK |
| 1828 | Fridge seems stable at 32.6 mK. Leaving 80V bias overnight to see if phonons return. |