2015/08/11 (Tuesday)

previous day

Overnight status: bath full of LN2, He exchange gas in IVC. MC fell to 80 K at about 20:30. Cryocleaning mixture overnight.
0905 MC = 77.4 K, increasing at about 0.01 K/hour. Sorb reading 127 K, increasing at about 0.5 K/hour. IVC = 1.4 mbar.
0910 Pump out IVC with leak detector. LD doesn't like that pure He much - close LD valve, vent and pump a few times. Begin pumping out IVC with turbo pump.
0922 Fill LN2 cold trap Dewar
0940 IVC pumping = 1.9e-6 mbar, IVC = 2.2e-5 mbar
1115 IVC pumping = 1.2e-6 mbar, IVC = 1.4e-5 mbar.
1120 Leak detector on IVC with turbo valved out, signal = 180 pbarL/s, slow upward creep. Start blowing LN2 out of bath with He gas.
1205 LD signal steadily dropping, now 100 pbarL/s.
1220 Cease cryocleaning and pump to dump.
1300 Perform LN temperature electrical check, all looks normal.
1305 After pumping mix to dump G2 reads 792 mbar.
1311 Conduct a throughput test , and get a throughput rate of 60 mbar/min in G1. (G1 allowed to rise to about 624 mbar and fell to minimum of 201 mbar.) Background leak rate remained unchanged from 100 pbarL/s during test, indicating no leak. Pump mixture back to dump. Switch from leak checker to turbo pump on IVC.
1342 Flush main bath with He gas. Leak rate continues to slightly decrease, currently at 91 pbarL/s.
1355 Commence LHe transfer (-1.0%).
1410 Cease pumping on IVC. Pressure w/o pumping is 5.4E-6mbar. Add 4.6E-2mbar of He exchange gas. Isolate IVC Leave IVC open to gauge.
1500 MC = 70 K, Sorb = 102 K, IVC = 3.4e-2 mbar.
1535 Sorb flatlines at 19.9 K
1600 MC = 41 K, Sorb = 19.0 K, IVC = 1.6e-2 mbar. Still = 22 K, Cold plate = 34 K, Tower = 44 K.
1630 MC = 27 K, Sorb = 7 K, IVC = 1.0e-2 mbar. 1K Pot = 5.8 K, Tower = 28 K.
1640 Main bath reading 4.8%
1630 MC = 16 K, Sorb = 5.5 K, IVC = 6.3e-3 mbar. 1K Pot = 4.6 K, Main bath 23.4%.
1700 MC = 12.5 K, Sorb = 5.4 K, IVC = 6.3e-3 mbar. 1K Pot = 4.5 K, Main bath 75.5%.
1707 Tower 14.2 K
1710 End LHe transfer (92.1%.). Tower has reached 11K, so with IVC pressure at 1.4E-3 mbar, commence pumping on IVC with turbo pump overnight.
1720 Sorb = 5.8 K, 1K Pot = 4.9 K, Still = 5.3 K, Cold plate = 8.3 K, Tower = 10.6 K, MC(Cernox) = 9.8 K, MC(H35) = 9.5 K, IVC pumping = 3.4e-6 mbar
1738 Fill LN2 cold trap full.

next day