UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20150723

2015/07/23 (Thursday)

previous day

0940 Fill LN2 cold trap full.
1005 Stop baking detectors.
1045 Tower Temp at 31.3mK. Commence phonon pulse study. Remove all but one DCRC board (DCRC33 on DIB2). The following table shows pulse shapes for various DCRC/filter combos when only DIB2 has a DCRC plugged into it. Ethernet cable #2 was used for all combinations. For pulse events, trigger thresholds set to values in table below:
pA 25
pB 15
pC 25
pD 25
1145 Unpower, add filter, and repower DCRC33 on DIB2. Set up the board and increase the QET biases by 20uA on every channel in order to see proper pulses.
DIB2 DCRC33 No Filter
DIB2 DCRC33 Filter
DIB2 DCRC33 No Filter Noise
DIB2 DCRC33 Filter Noise
1216 Unpower and remove DCRC33 from DIB2. Attach DCRC50 with no filter to DIB2. Find and set SQUID Bias and Lockpoints, as shown in table below. Estimate QET Biases by eye.
DIB2 DCRC50 SQUID Settings
Channel Bias (mA) Lockpoint (mV) Vpp (mV) QET Bias (mA)
A -76.17 -4.18 317 90
B -64.45 -1.46 270 90
C -78.12 -2.72 320 110 or 90?
D -78.12 -2.63 337 90
1235 Take pulse and noise pictures for DCRC50 in DIB2 with no filter. Use trigger thresholds listed below.
pA 20
pB 15
pC 25
pD 25
1245 Unpower, add filter, and repower DCRC50 on DIB2. Set up the board and increase the QET biases by 20uA on every channel in order to see proper pulses.
DIB2 DCRC50 No Filter
DIB2 DCRC50 Filter
DIB2 DCRC50 No Filter Noise
DIB2 DCRC50 Filter Noise
1316 Unpower and remove DCRC50 from DIB2. Attach DCRC55 with no filter to DIB2. Find and set SQUID Bias and Lockpoints, as shown in table below. Estimate QET Biases by eye.
DIB2 DCRC55 SQUID Settings
Channel Bias (mA) Lockpoint (mV) Vpp (mV) QET Bias (mA)
A -76.17 -2.82 330 100
B -64.45 -1.17 259 100
C -72.27 -3.31 284 120
D -80.08 -2.04 269 100
1320 Notice that DIB2 Channel B has weird modulation curves for both DCRC50 and DCRC55. Picture of modulation curve included below for clarity.
DIB2 DCRC55 Phonon Channel B Modulation Curve
1340 Take pulse and noise pictures for DCRC55 in DIB2 with no filter. Use trigger thresholds listed below.
pA 20
pB 15
pC 15
pD 25
1350 Notice odd noise sometimes present in DCRC55 pulses with no filter in. Picture included for clarity.
DIB2 DCRC55 Odd Noise in Phonon Pulses
1410 Unpower, add filter, and repower DCRC55 on DIB2. Set up the board and increase the QET biases by 20uA on every channel in order to see proper pulses.
DIB2 DCRC55 No Filter
DIB2 DCRC55 Filter
DIB2 DCRC55 No Filter Noise
DIB2 DCRC55 Filter Noise
1536 Unpower and remove DCRC55 from DIB2. Attach DCRC31 with no filter to DIB2. Find and set SQUID Bias and Lockpoints, as shown in table below. Estimate QET Biases by eye.
DIB2 DCRC31 SQUID Settings
Channel Bias (mA) Lockpoint (mV) Vpp (mV) QET Bias (mA)
A -76.17 -2.82 330 100
B -64.45 -1.17 259 100
C -72.27 -3.31 284 120
D -80.08 -2.04 269 100
1545 Take pulse and noise pictures for DCRC31 in DIB2 with no filter. Use trigger thresholds listed below.
pA 20
pB 15
pC 15
pD 20
1615 Unpower, add filter, and repower DCRC31 on DIB2. Set up the board and increase the QET biases by 20uA on every channel in order to see proper pulses.
DIB2 DCRC31 No Filter
DIB2 DCRC31 Filter
DIB2 DCRC31 No Filter Noise
DIB2 DCRC31 Filter Noise

next day