start › cdms › k100 › run_diary › 20150703

2015/07/03 (Friday)

previous day

status Flashing overnight, Cf source at detector level. DIB1/2/3 → Filter none/100MHz/3.2MHz → DCRC55/33/50.
0950 Tower = 298 mK, Still = 2 mW / 0.13 mbar, G1/G2 = 260/240 mbar, 1K pot = 1.69 K / 14 mbar, Main bath = 37.9%
0954 turn off flashing
1001 fill LN2 cold trap Dewar
1052 Trying to set DCRCs. 3KHz 5mVpp oscillation in DIB2qI. Unplug DCRC33.
1105 Tower base falls from 32.1 to 31.5 mK with DCRC33 off. Repower DCRC33.
1123 Zap DIB2 SQUIDs.
1140 remove Cf source
Current FrontendGain/OutputGain(setting)/DriverOffset settings
pA pB pC pD
DIB1 7/1/0.170 7/1/0.156 7/1/-0.014 7/1/0.074
DIB2 7/1/0.022 4/1/0.036 7/1/0.028 7/1/0.022
DIB3 7/1/-0.074 7/1/-0.012 7/1/0.198 7/1/0.172
1152 DIB1 pA and pC keep losing lock. Oscillation in DIB1 phonons, 150 Hz, 5/10/20/10 mVpp in pA/pB/pC/pD.
1202 try zapping DIB1 SQUIDs
1218 Now all four DIB1 phonon pulses upside down :(. 150 Hz oscillation seen in all three DIBs, stronges in DIB1pA (15 mVpp) and DIB2 pC (20 mV). Unplug DCRC55 from DIB1.
1239 Repower DCRC55/DIB1, zap SQUIDs, set timing link
1401 new driver offsets are closer to “normal”:
Updated DriverOffset settings
pA pB pC pD
DIB1 -0.092 -0.082 -0.028 -0.094
1402 1 kHz noise in DIB1 phonon channels
1417 Begin LHe transfer at 31.5%
1455 End LHe transfer at 93.7%
1515 Ugly square wave in DIB1 phonons, 150 Hz, 5/0/10/4 mVpp.
1516 Set optimal phonon thresholds:
Current phonon threshold settings
pA pB pC pD
DIB1 30 25 35 20
DIB2 25 20 25 30
DIB3 15 15 10 20
1520 flash 30 seconds
1532 Set to -4/+4 V, start bkg data
1549 Stop data, set to 0/0 V
1555 Place Cf source at detector level
1558 Looks like a lot of channels railed, probably during run. Oscillation on DIB1qI, elevated temperature (was 31 mK, now 32 mK). Looks like DCRCs are in a Bad State. May as well unplug DCRCs.
1620 Fill LN2 cold trap Dewar