2015/06/17 (Wednesday)

0849 Fill LN cold trap Dewar
0920 Currently there is an oscillation in the phonon channels (seen in DIB2 but likely present in the others too), low frequency: period is about 5 ms (so it's more clearly seen with the pulse window set to 20 ms). It is varying in amplitude and there are spikes in Tower temperature every few minutes, between around 2 and 20 mK.
0922 MC/Tower = 26.3/33 mK, Cold plate = 146 mK, Still = 1.11 K / 0.080 mbar, G1/G2 = 175/165 mbar, 1K Pot = 1.70 K / 18 mbar, Main bath = 14.5%.
0934 Problem with low-frequency oscillation seen in all phonon channels. Period is about 5 ms, about 40/20 mVpp in DIB1/2 with both gains set to 1. Also there are tower temperature spikes every few minutes around 2-20 mK high.
overnight temperature was stable, until starting at 6:10 temperature spikes were seen at 20 minute intervals repeating four times (last at 7:20), each a little stronger than the last, from about 1 mK to 5 mK in height. Then stable till about 8:30 when Anthony came in and began doing stuff with DCRCs.
0938 DIB1 Qouter also shows a lowish amplitude oscillation, period around 1 ms.
0939 attempted power cycling each DCRC one at a time, did not remove oscillation.
0950 Turning off DIB1 FETs seems related - at one point leaving them off for a while seemed to make oscillation go away. But as we sometimes saw in the previous run, turning off the FETs with DCRC31 often caused the temperature to increase.
1123 Disconnected power to all DCRCs. Strangely there were additional temperature spikes:
1136 Tower spiked to 63 mK
1145 MC spike from 28 to 38 mK
1156 Tower spike from 33 to 75 mK, MC spike from 27 to 38 mK, MC spike seems delayed rel. to Tower. Some heat coming from Tower for some reason…?
1227 No more temperature spikes. MC/Tower = 25/32 mK. Still pressure is rather low, around 0.075 mbar, so try apply 2 mW to still. Still P increases to 0.100 mbar, MC stabilizes a little colder, 24 mK.
1334 MC/Tower stable at 24/31 mK (even lower base for tower!).
1411 Commence LHe transfer (5.0%).
1446 Cease LHe transfer (64.3%).
1655 Commence retaking SQUID lockpoint sweep data taking after a 20 minutes flash and 2 hour cooldown: 7 data series each of length 10 minutes, taken at -/+4.0 V bias, interspersed by 6-second flash and 20-minute cooldown periods. OR trigger on phonon channels each with a threshold of 50 ADC Bins. Front End Gain of 7, Output Gain of 1. Squid Lockpoints used included in table below:
SQUID Lockpoints [mV]
Series DIB1 A DIB1 B DIB1 C DIB1 D DIB2 A DIB2 B DIB2 C DIB2 D DIB3 A DIB3 B DIB3 C DIB3 D
1 5.91 6.3 4.06 3 5.13 2.9 3.38 2.9 -4.76 -2.91 -2.04 -2.62
2 5.95 6.34 4.1 3.04 5.17 2.94 3.42 2.94 -4.73 -2.88 -2.01 -2.59
3 5.99 6.38 4.14 3.08 5.21 2.98 3.46 2.98 -4.7 -2.85 -1.98 -2.56
4 (Nominal Values) 6.03 6.42 4.18 3.12 5.25 3.02 3.5 3.02 -4.67 -2.82 -1.95 -2.53
5 6.07 6.46 4.22 3.16 5.29 3.06 3.54 3.06 -4.64 -2.79 -1.92 -2.5
6 6.11 6.5 4.26 3.2 5.33 3.1 3.58 3.1 -4.61 -2.76 -1.89 -2.47
7 6.15 6.54 4.3 3.24 5.37 3.14 3.62 3.14 -4.58 -2.73 -1.86 -2.44