2015/06/15 (Monday)

0850 MC/Tower = 25.9/31.9 mK, Cold plate = 145 mK, Still = 1.11 K / 0.090 mbar, G1/G2 = 176/166 mbar, 1K Pot = 1.69 K / 17 mbar, Main bath = 11.6%.
0900 Fill LN cold trap Dewar
1215 Finish LHe transfer
1800 Fill LN cold trap Dewar
1830 Commence SQUID lockpoint sweep data taking after a 20 minutes flash and 2 hour cooldown: 7 data series each of length 10 minutes, taken at -/+4.0 V bias, interspersed by 6-second flash and 20-minute cooldown periods. OR trigger on phonon channels each with a threshold of 50 ADC Bins. Front End Gain of 7, Output Gain of 1. Squid Lockpoints used included in table below:
SQUID Lockpoints [mV]
Series DIB1 A DIB1 B DIB1 C DIB1 D DIB2 A DIB2 B DIB2 C DIB2 D DIB3 A DIB3 B DIB3 C DIB3 D
1 5.91 6.3 4.06 3 5.13 2.9 3.38 2.9 -4.76 -2.91 -2.04 -2.62
2 5.95 6.34 4.1 3.04 5.17 2.94 3.42 2.94 -4.73 -2.88 -2.01 -2.59
3 5.99 6.38 4.14 3.08 5.21 2.98 3.46 2.98 -4.7 -2.85 -1.98 -2.56
4 (Nominal Values) 6.03 6.42 4.18 3.12 5.25 3.02 3.5 3.02 -4.67 -2.82 -1.95 -2.53
5 6.07 6.46 4.22 3.16 5.29 3.06 3.54 3.06 -4.64 -2.79 -1.92 -2.5
6 6.11 6.5 4.26 3.2 5.33 3.1 3.58 3.1 -4.61 -2.76 -1.89 -2.47
7 6.15 6.54 4.3 3.24 5.37 3.14 3.62 3.14 -4.58 -2.73 -1.86 -2.44