start › cdms › k100 › run_diary › 20150605

2015/06/05 (Friday)

MC & Sorb temperatures slowly decreased overnight.
0817 MC(Cernox) = 11.3 K, Tower = 12.6 K, Cold plate = 18.1 K, Still = 9.3 K, 1K Pot > 7 K, Sorb = 13.7 K. Main bath = 56.8%.
0823 IVC pumping overnight. Pumping pressure = 1.6e-6 mbar, IVC pressure = 1.0e-5 mbar
0825 Room temperature LN traps pumping out overnight, P2 = 0.013 mbar
0830 Put LN traps back in Dewar and fill Dewar
0835 v9 open, G2 = 795.4 mbar, barometric pressure = 30.16 inHg, corrected dump pressure = 795.4+33.86*(30.19-29.92) = 804.5 mbar
0840 Pump GHS volumes to dump through v5, v1, v3. Then close these and pump through v6 to dump.
0855 P1 = 0.007 mbar, G1 = 1.0 mbar. G2 = 796.0 mbar, barometric pressure = 30.16 inHg, corrected dump pressure = 805.1 mbar
0858 Open V13a to let mix into dump, allow to cool
0917 Open condenser valve, v1, and cold trap 1 speedivalve to let mix into fridge.
When mix was let in, MC warmed slightly and Sorb and 1K Pot cooled significantly.
0929 MC(Cernox) = 10.9 K, Tower = 12.1 K, Cold plate = 17.3 K, Still = 7.5 K, 1K Pot = 6.6 K, Sorb = 7.7 K.
0951 G1 & G2 equalized at 620 mbar
0957 MC(Cernox) = 10.7 K, Tower = 11.9 K, Cold plate = 16.4 K, Still = 6.4 K, 1K Pot = 5.9 K (flat), Sorb = 7.0 K (flat).
1000 Perform electrical checks at breakout box.
1018 Before starting 1K Pot, close V13a (keeping most mix in dump)
1022 Open 1K Pot needle valve 60 degrees. Quick drop of 1K Pot temperature to 4.4 K, Sorb to 5.3 K
1027 Open 1K Pot speedivalve
1039 1K Pot = 1.69 K / 17 mbar. G1 = 150 mbar. (G2 = 614 mbar, left in dump.) MC(Cernox) = 7.8 K, Tower = 8.4 K, Cold Plate = 3.3 K, Still = 2.1 K , Sorb = 4.0 K flatline (I guess this is minimum).
1044 Open V13a to let mix from dump into fridge (12a still closed, but trap1 speedivalve open - trickling mix in)
1101 Mix entering and condensing, G1/G2 = 83/512 mbar, both falling. MC(H35) = 2.25 K, Tower = 2.26 K, Cold plate = 2.77 K, Still = 1.62 K.
1101 1K Pot = 1.63 K / 8.5 mbar - lowish. Open needle valve additional 10 degrees.
1107 1K Pot = 1.79 K / 55 mbar
1108 Open V12a to 0.3% to let mix in more quickly
1131 G1/G2 = 75/280 mbar and falling. MC(H35) = 1.92 K, Tower = 1.93 K, Cold plate = 2.30 K, Still = 1.84 K, 1K Pot = 1.79 K / 50 mbar.
1137 Open V12a to 5% to let mix in more quickly
1244 G1/G2 = 50/50 mbar. MC(H35) = 1.85 K, Tower = 1.86 K, Cold plate = 2.21 K, Still = 1.77 K, 1K Pot = 1.78 K / 45 mbar.
1255 Open V12a completely.
1300 Close V9, leaving 44 mbar in dump.
1301 Open gate valve to still pumping line. This causes a surge in 1K Pot pressure & temperature - interesting. Perhaps it's just more warm mix being pulled in through the condenser side as the pressure is released from the still side into the still pumping volume.
1307 Start circulation procedure. V6 open at 12.0%.
1314 No change in pressures until V6 opened to 13.0%.
1315 MC(H35) = 1.86 K, Tower = 1.87 K, Cold plate = 2.19 K, Still = 1.82 K, 1K Pot = 1.78 K / 45 mbar.
1337 At V6 = 13.8%, P1 pressure begins to fall into readable range. G1/G2 = 160/190 mbar, Still = 1.44 K / 85 mbar. MC and Cold plate down to 1.4 K.
1349 V6 = 15.0%, G1/G2 = 125/130 mbar, Still = 1.04 K / 2.8 mbar. MC/Tower = 0.82/0.80 K, Cold plate = 0.75 K.
1357 V6 = 16.0%, G1/G2 = 136/136 mbar, Still = 1.05 K / 0.85 mbar. MC/Tower = 680/650 mK, Cold plate = 556 mK.
1401 V6 = 18.0%, G1/G2 = 124/124 mbar, Still = 1.07 K / 0.38 mbar. MC/Tower = 285/274 mK, Cold plate = 470 mK.
1405 Open V6 to 100%
1411 Open V14 to pump the rest of the mix into the fridge
1422 P1 down to 3.8 mbar, dump mostly empty - close V14
1423 G1/G2 = 76/76 mbar, Still = 1.05 K / 0.12 mbar. MC/Tower = 59/64 mK, Cold plate = 147 mK. ←- good! Cold plate was stuck at 300mK last run!
1427 1K Pot = 1.79 K / 50 mbar. Close needle valve a little.
1436 MC/Tower = 34/41 mK (already better than last run!) Cold plate = 138 mK, Still = 1.04 K / 0.12 mbar, 1K Pot = 1.74 K / 32 mbar.
1449 MC/Tower = 28.3/37.0 mK. Cold plate = 137 mK, Still = 1.05 K / 0.11 mbar, 1K Pot = 1.74 K / 30 mbar. G1/G2 = 84/81 mbar.
1510 MC/Tower = 26.6/35.8 mK. Cold plate = 137 mK, Still = 1.05 K / 0.11 mbar, 1K Pot = 1.74 K / 30 mbar. G1/G2 = 98/92 mbar.
1600 MC/Tower = 26.4/35.6 mK. G1/G2 = 125/116 mbar.
1625 Commence LHe transfer (37.6%).
1704 End LHe transfer (93.6%).
1725 Fill LN2 cold trap full.
1726 MC/Tower = 26.2/35.5 mK. G1/G2 = 120/113 mbar. The pressures are creeping up - hopefully this will not be a problem.