start › cdms › k100 › run_diary › 20150520

2015/05/20 (Wednesday)

0853 MC/Tower = 52.0/66.6 mK, G1/G2 237/214 mbar, Still = 4mW / 1.08K / 0.16 mbar, 1K Pot = 1.73 K / 22 mbar, Cold plate = 9.47 kΩ / 307 mK, MC sensor = -39.2 kΩ / 68.8 mK, Main bath = 36.6%
0904 Fill LN cold trap Dewar
0914 One-shot test to study the cryogenic situation: monitor MC, Tower and Cold Plate temperatures. Observations:
1 MC temperature fell to near 30 mK in 10 minutes, then began to rise again. Tower temperature followed along with Tower-MC increasing from 15 mK to about 20 mK.
2 G2 pressure increases linearly as mix is collected in dump. Temperature minimum occurs at G2 = 77 mbar. Stop one-shot when G2 = 100 mbar.
3 Cold plate initially gets warmer. After about 5 minutes it begins to cool along with the MC. CP-MC starts at about 215 mK, increases to 234 mK, then stays roughly constant.
4 Still gets colder. Still pressure falls by roughly 1/2. (Still heat was left at 4 mW - likely if it were heated more during the one-shot an even colder minimum could be achieved.)
#3 above has interesting implications in terms of heat leaks. There are four main paths of heat from the CP stage to the MC stage: through the heat exchanger, through the tower, IR radiation, and through the tower thermometer wire assuming there is a touch. Since the CP got initially warmer, one of these must have been “disconnected” - of course that's the heat exchanger path, because returning mixture is what makes the thermal connection between MC and CP at the heat exchangers. But when CP-MC gets large enough, one or more of the other heat paths becomes strong enough to cause further cooling.
0934 Restore circulation, with 100 mbar of mixture in dump. MC gets warmer, to 196 mK.
1006 Move all mixture from dump into circulation. MC cools to 61 mK, seems to get stuck there - try increasing still heat from 4 mW to 8 mW. MC cools back to previous base, 53 mK.
1025 MC/Tower = 53/67 mK, G1/G2 = 282/256 mbar, Still = 8 mW/ 1.09 K / 0.20 mbar, 1K Pot = 1.70 K / 16 mbar, Cold plate = 9.51 kΩ / 306 mK. (note: 1K Pot needle valve was closed a bit at one point. It could be that with a lower pot temperature you want a higher still temperature for optimal cooling.)
1051 Tried connecting 3 DCRCs, DIB 1/2/3 = DCRC 31/50/33. But the interference problem returned repeatedly - could get rid of it by power cycling the DIB1 FETs but only for a couple of seconds.
1130 Connect DCRC31 → DIB1, DCRC33 → DIB2.
1212 All SQUIDs zapped and set, QET biases set, timing link set. Try to find a stable crystal bias… goal is nominally -4V/+4V, so 8V total - could be asymmetric but probably want some non-zero bias on either side, otherwise a lot of charge will be collected on guard rails (highly reduced charge collection efficiency for that side).
1224 Try many biases to test for stability, defined as how long it will last before slipping into the interference state (hopefully indefinitely). Side 1/2 = 0V/+3V, -2V/0V, and 0V/+8V all seemed stable. Tried -0.5V on Side 1 and various settings on Side 2… anything above +4.5 V on Side 2 seemed to only be stable for a few minutes. -0.5V/4.5V may have been stable, at least lasted for 13 minutes one time.
1520 With charge biases at zero (to avoid interference) tried setting the QET gains all to Front End = 50, Output = 7. This generally worked - in all cases I could find an appropriate Driver Offset so that pulses appeared in the window (this is easier to do with the y-axes range constant, e.g. -1.3 V to +1.3 V).
notes: with DIB2 pC, any Front End Gain greater than 7 caused high-frequency noise in all the other channels from that DIB. With DCRC31, I am pretty sure that changing Front End Gain has no effect. This is based on comparing it to DIB2/DCRC33, where changes in the QET Front-End Gains always caused noticeable changes in the phonon channel displays, e.g. change in the baseline offset position.
more evidence that the DCRC31 Front-End Gains aren't working: if you trigger on a phonon channel, there are certain threshold settings (around 500 to 800 ADC units) where there is a marked difference in event rate if you modify either gain setting significantly - this is seen in DCRC33 and in the Output Gain for DCRC31, but it is not seen in the Front-End Gain for DCRC31.
1602 Main bath = 26.2%. Hopefully it will last until tomorrow morning.