MC dropped to <3 K and flat-ish at around midnight, with G1 dropping to 100 mbar and flat. Condensation finally.
0831 MC 2.23 K (Cernox), Tower 2.25 K, Still 1.82 K. 1K Pot 1.78 K / 45 mbar. Main bath 53%. G1/G2 89/304 mbar.
0840 fill LN trap dewar
0845 open V13a for about 40 minutes to let more mix in. Left 150 mbar in dump.
0922 finish electrical check. Compared these 2K results to 80K results: in DIB1, the shorts between Qi/Qo and the QET returns seem to have disappeared, and LED2 seems to be open now. In DIB2, the shorts across QET A (bias to return) and QET B have disappeared, but both LED1 and LED2 appear to be open now.
0934 Open V12a full. MC = 1.92 K (H35), 1.95 K (Cernox). Tower = 1.93 K. G1/G2 = 56/56 mbar (150 mbar still in dump). Still = 1.83 K. 1K Pot = 1.76 K / 38 mbar.
0940 Begin circulation
0945 V6 open to 13.2%, G1/G2 = 96/113 mbar and stabilizing, Still = 1.69 K and falling.
0954 P1 falls to 40 mbar. V6 at 13.6%, G1/G2 = 115/113 mbar and falling, Still = 1.3 K
During initial circulation, Still falls to 1.10 K, then wiggles between 1.04 and 1.10 K.
1018 V6 fully open, G1/G2 = 113/120 mbar and stable. Still = 1.04 K / 0.220 mbar. 1K Pot = 1.8 K / 40 mbar.
1036 MC falls to 40 mK, Tower to 68 mK
1040 Still 1.04 K / 0.140 mbar, 1K Pot 1.76 K / 40 mbar
1043 Open V14 briefly to introduce more mix from dump, 2 times in 5 minutes. Still cools to 0.96 K during process.
1050 Open V14 and leave it open for 8 minutes. Closed V14 when P1 fell to 3.6 mbar.
1101 Still 1.04 K / 0.140 mbar, G1/G2 = 82/89 mbar, Tower 92 mK and falling. MC falls from 68 to 52 mK with +/- 5 mK oscillations
1107 Close needle valve a bit on 1K Pot, reducing to 1.69 K / 18 mbar
1126 Tower 76 mK, MC 52 mK.
1230 MC has been oscillating between 52 and 62 mK in a shark-fin pattern, about 6 minute period. Pumping pressures oscillate in sync (not surprisingly). Tower much warmer, oscillating between 73 and 77 mK.
1320 MC and Tower temperature still oscillating. Commence heating still to increase circulation rate.
1335 Heating still to 4 mW causes oscillations to cease. P1 = 0.18 mbar, MC = 53 mK, Tower = 73 mK
1346 Still = 6 mW, 0.20 mbar. Base rises then falls again, settling at same base, MC/Tower = 53/73 mK
1408 Still = 6 mW, 1.08 K, 0.20 mbar. G1/G2 = 84/95 mbar. 1K Pot = 1.69 K, 17 mbar. Coldplate = 0.31 K.
1421 Switch to one-shot mode. Base temperatures fall quickly. After 7 minutes, MC/Tower = 30/59 mK. After another 4 minutes, MC/Tower = 27/56 mK.
1435 When MC falls to 26 mK and 80 mbar recovered to dump, switch back to circulation mode. (note still heater has been left at 6 mW throughout.) Base temperatures rising again.
1458 MC back to 53 or 54 mK
1500 Reduce still heater to 4mW, P1: 0.18mbar. No change in base temperature; MC temperature still at 54mK. G1/G2: 75/88mbar.
1525 Open needle valve to raise 1K Pot temperature to 1.656K. P2: 9.5mbar.
1600 Commence LHe transfer (36.7%)
1605 After oscillating ~53.5mK+/-0.3mK, raise 1K Pot temperature to 1.695K.
1640 MC temperature begins oscillating ~53.5mK+/-0.5mK. Not better.
1645 Turn down 1K Pot in attempt to stop oscillations.
1655 Commence LHe transfer (76.0%)
1715 Fill LN2 cold trap full.
1750 Set IP addresses of DCRC 33, DCRC 50. Unable to set IP adress of DCRC 31. Need new driver for PC (possibly). Emailing Steen.