| 0945 | Set up mini clean room |
| 1015 | Found a bad connection on the tower thermometer plug, so redid the soldering. Used a new Samtec(?) connector, since the other one had silicone on it. The connector used had posts without holes - I wrapped the wire around a few times before soldering, hopefully this is secure and stress-relieving. Taped it over with Kapton tape. |
| | actual part used was 3M Headers & Wire Housings FULL STICK HDR/36POS - http://www.mouser.com/ProductDetail/3M-Electronic-Solutions-Division/929850-01-36-RA/?qs=VCIs6OJqWMmA2mkHtYsWPQ%3D%3D |
| 1245 | checked striplines from cold end to ebox for shorts to ground. No shorts on any pins. |
| 1300 | Installed tower |
| 1700 | During electrical checkout after attaching SQET cards, found a bad connection in Stripline 19 - the QET Bias B wire. We can probably fix this with wire-rerouting cards at SQET end and E-box end. |