0850 Dump reads 794.1 mbar at G2. Pbar = 30.12“Hg, corrected dump pressure = 800.9 mbar
Main bath 61.4%. MC 7.1 K on Cernox thermometer, peaked at 7.3 K overnight
Turbo on IVC reads 1.9e-6 mbar. IVC reads 1.4e-5 mbar with turbo valved out
0920 Warm up cold traps and pump them out. T1 had 40.0 mbar, T2 had 8.0 mbar, both read at P2. Pumped each down to 0.024 mbar
0950 Letting mixture into fridge before starting 1K Pot (to look at temperatures)
G2 (dump) starts at 795. Trickle mix into fridge through trap and closed V12A
Before starting, MC (H35) = 7.1 K, Still (U02844) = 9.1 K
When mix goes in, MC heats to 7.35 K and flattens. Still increases to 9.4 K but then steadily falls
this suggests that overnight the still gets somewhat warmer than the bath due to conduction from the top. When mix is introduced it is cooled by the bath above the IVC, and the still gets cooler as it approaches equilibrium with the bath.
1017 Still (U02844) = 7.7 K
1022 Still (U02844) = 7.2 K, below MC which is at 7.35 K. 1K Pot temperature falls to 6.5 K
1025 G2=602 mbar (dump slowly emptying), G1 = 385 mbar (mix entering fridge)
1030 Still (U02844) = 6.73 K, MC = 7.28 K (slowly cooling now), 1K Pot = 6.27 K
1037 Still (U02844) = 6.38 K, MC = 7.20 K, 1K Pot = 6.09 K. G2 = 578 mbar, G1 = 451 mbar
1051 Still (U02844) = 6.00 K, MC = 7.05 K, 1K Pot = 5.90 K. G2 = 578 mbar, G1 = 451 mbar
1103 Still (U02844) = 5.85 K, MC = 6.93 K, 1K Pot = 5.84 K. G2 = 552 mbar, G1 = 522 mbar
1107 Start 1K Pot, needle valve open slightly, 4.0 mbar. 1K Pot falls from 5.8 K to 3.1 K, Still falls from 5.8 K to 3.3 K, G2 falls from 530 mbar to 440 mbar.
1124 Open needle valve in small steps. 1K Pot at 2.2 K / 8.0 mbar. MC plummets to 3.0 K. Still 2.6 K and falling. G1 235 mbar and falling. G2 (dump) 522 mbar, slowly falling.
1133 Open V12a to 2%. Temperatures go up a bit, G2 starts rising, G1 falls faster.
1139 G2 = 414 mbar, G1 = 174 mbar
1147 Still (U02844) = 2.53 K, MC = 2.55 K, 1K Pot = 2.05 K / 12 mbar. G2 = 321 mbar, G1 = 181 mbar and falling again
1155 Still (U02844) = 2.15 K, MC = 2.30 K, 1K Pot = 1.66 K / 12 mbar. G2 = 249 mbar, G1 = 106
1206 Still (U02844) = 1.88 K, MC = 2.02 K, 1K Pot = 1.62 K / 8.5 mbar. G2 = 165 mbar, G1 = 65
1230 open V12A full
1236 forgot to isolate IVC from turbo pump, doing so now
1245 Still (U02844) = 1.61 K, MC = 1.66 K, 1K Pot = 1.55 K / 5.0 mbar. G2 = 34 mbar, G1 = 34 mbar (note we have now let in almost all the mix)
1248 Begin circulation
tried the standard throttling of V6 to control G1/G2 but had trouble, oscillatory behavior in still temp and pressures, hard to get P1 to come down except very briefly
after a while opened V6 wide, and finally saw still temperature fall below 1.3 K and fall to 0.75 K. 1K pot had trouble at first, pressure at P2 maxed out, but after a few minutes everything looked good.
1343 MC = 0.218 K, Still (U02844) = 0.846 K, 1K Pot = 1.56 K / 5.0 mbar. G1 = 86 mbar, G2 = 93 mbar
1351 MC = 0.112 K, Still (U02844) = 0.956 K / 0.10 mbar, 1K Pot = 1.59 K / 6.0 mbar. G1 = 84 mbar, G2 = 80 mbar
1357 MC = 60.0 mK, Still (U02844) = 0.950 K / 0.085 mbar, 1K Pot = 1.65 K / 12 mbar. G1 = 87 mbar, G2 = 84 mbar
1404 MC = 39.2 mK, Still (U02844) = 0.962 K / 0.065 mbar, 1K Pot = 1.64 K / 9.5 mbar. G1 = 77 mbar, G2 = 75 mbar
1515 G1=71 mbar, G2=72 mbar, P1=60 ubar, and mixing chamber temperature at 29 mK and stable. Open V14 to bleed out remaining mix from dump. P1 rose to 36 mbar and begun to drop.
1537 Close V14, with mixing chamber temperature at 33mK.
1633 MC = 28.3 mK, Still (U02844) = 0.951 K / 0.085 mbar, 1K Pot = 1.65 K / 12.0 mbar. G1 = 67 mbar, G2 = 66 mbar. Fridge quite stable.
Applied still heat for the purpose of recalibrating the still thermometer. As with the last run, heating the still did not improve the base temperature.
1800 Recovering mix to dump.
2040 Recovered 799 mbar dump pressure of mixture. Turn off all heaters and close all valves - stop mixture recovery.
Pbar = 30.02”Hg, corrected dump pressure = 802.4 mbar