0857 Dump measured at G1: 795.4 mbar
Took stripline tube to shop to have them remove the scratches.
Attached the U02844 thermometer (tower thermometer) to the still plate. Purpose is to calibrate still thermometer.
Heat sunk wire from thermometer to post on still plate, then more thoroughly to post on 100 mK can, then ran it to plug at MC stage
Also covered hole in 100 mK can top with foil, to block radiation from 600 mK/4 K to top of MC
At room temperature U02844 reads 1.00 Ohm at the LakeShore bridge. This seems about right according to calibration curve.
1545 Perform Stripline Electrical check. Results below:
Stripline Bad Pins
19 No bad pins. No shorts to ground.
20 Qo Bias (FET Pin 21). No shorts to ground.
27 Feedback coil B, Feedback coil C, QET D Return. No shorts to ground.