Summary of mixture amounts:

morning of G2 (mbar) Pbar (“Hg) Pdump (mbar) 3He (mbar) 3He (%)
Mar 23 653 30.3 665.9 166.5 25%
Mar 24 700 30.1 706.1 166.5 23.6%
Mar 26 746.3 30.1 752.4 166.5 22.1%
Mar 27 789.5 30.3 802.4 166.5 20.8%

The barometric pressure (Pbar) is from Weather Underground (wunderground.com) weather history for Minneapolis. This is Sea Level Pressure (so not the actual measured pressure but rather corrected to remove the elevation of the weather station). The original 25% is based on the mass spec measurement and has a +/- 3% error. The other percents are inferred from this original measurement.

0900 fill LN trap dewar
0918 pump fridge contents to dump. Dump reads 789.5 mbar at G2.
Pbar = 30.3 “Hg, so corrected for barometric pressure, Dump = 802.4 mbar
0948 Start 1KPot
1000 Begin condensing mixture
1043 Condensing, V12a open 100%. G1/G2 = 72/74 mbar, MC = 2.1 K
1050 Begin circulation. V6 open 13%
1113 V6 = 14.2%, MC/Still = 1.13/1.17 K
1128 Open V6 100%
1130 MC = 0.87 K, Still = 0.87 K / 0.22 mbar
1131 MC 776 mK, Still 921 mK / 0.20 mbar, G1/G2 131/124 mbar, 1KPot 1.77 K / 45 mbar
1139 Open V14
1144 MC 593 mK, Still 896 mK / 5.0 mbar, G1/G2 106/107 mbar, 1KPot 1.77 K / 45 mbar
1157 MC 526 mK, Still 948 mK / 1.3 mbar, G1/G2 133/127 mbar, 1KPot 1.77 K / 45 mbar
1208 MC 474 mK, Still 986 mK / 0.65 mbar, G1/G2 134/127 mbar, 1KPot 1.76 K / 45 mbar
1210 With P1=0.6 mbar, close V14.
1216 MC 448 mK, Still 992 mK / 0.20 mbar, G1/G2 135/128 mbar, 1KPot 1.76 K / 45 mbar.
1241 MC 471 mK, Still 951 mK / 0.17 mbar, G1/G2 121/117 mbar, 1KPot 1.70 K / 22 mbar.
1303 MC 475 mK, Still 951 mK / 0.17 mbar, G1/G2 119/116 mbar, 1KPot 1.70 K / 22 mbar.
1316 MC 477 mK, Still 951 mK / 0.17 mbar, G1/G2 122/119 mbar, 1KPot 1.70 K / 22 mbar.
1355 MC 477 mK, Still 951 mK / 0.19 mbar, G1/G2 127/123 mbar, 1KPot 1.70 K / 22 mbar.
1417 MC 409 mK, Still 1.05 K / 0.24 mbar, Still heater 5mW, G1/G2 136/135 mbar, 1KPot 1.70 K / 22 mbar.
1425 MC 342 mK, Still 1.11 K / 0.32 mbar, Still heater 10mW, G1/G2 141/146 mbar, 1KPot 1.70 K / 22 mbar.
1433 MC 321 mK, Still 1.17 K / 0.38 mbar, Still heater 15mW, G1/G2 151/157 mbar, 1KPot 1.70 K / 22 mbar.
1450 Begin a series of single-shots to remove 3He to dump. Removed 87.0 mbar, measured at G1. Left still heater on at 15 mW. Still stayed below 1.2 K and MC below 450 mK during procedure. Restored circulation with reduced mix.
1513 MC 560 mK, Still 1.17 K / 0.34 mbar, Still heater 15mW, G1/G2 137/143 mbar, 1KPot 1.70 K / 22 mbar.
1517 MC 574 mK and relatively flat, begin applying more heat to still.
1525 MC 507 mK, Still 1.21 K / 0.38 mbar, Still heater 20mW, G1/G2 145/154 mbar, 1KPot 1.70 K / 22 mbar.
1532 MC 472 mK, Still 1.28 K / 0.40 mbar, Still heater 25mW, G1/G2 155/165 mbar, 1KPot 1.70 K / 22 mbar.
1541 MC 442 mK, Still 1.27 K / 0.50 mbar, Still heater 30mW, G1/G2 177/187 mbar, 1KPot 1.70 K / 22 mbar.
1549 MC 415 mK, Still 1.33 K / 0.60 mbar, Still heater 40mW, G1/G2 208/220 mbar, 1KPot 1.72 K / 28 mbar.
As before, with less 3He, we can still get dilution cooling, and improve MC temperature by increasing still heat. But both still and (more importantly) MC temperatures are higher.
1555 Dialing still heat back to 15 mW in preparation of putting all the mix back in. Adding mix to circulation through V14.
1629 Close V14 when P1 reads 0.9 mbar
Performing Cooling Power Test with all mix in and still heater at 15 mW
1636 MC heater = 0 uW, MC temp = 324.4 mK
1640 MC heater = 10 uW, MC temp = 327.4 mK
1645 MC heater = 20 uW, MC temp = 327.1 mK (wiggles between 327 and 328.5)
1650 MC heater = 30 uW, MC temp = 327.9 mK
1656 MC heater = 40 uW, MC temp = 330.0 mK (dip)
1708 MC heater = 50 uW, MC temp = 330.3 mK
1712 MC heater = 60 uW, MC temp = 331.9 mK (dip)
1716 MC heater = 70 uW, MC temp = 333.5 mK
1720 MC heater = 80 uW, MC temp = 332.7 mK (dip)
1723 MC heater = 90 uW, MC temp = 336.0 mK
1726 MC heater =100 uW, MC temp = 337.3 mK
The “dips” above are brief (~1 min) dips in the MC temperature, then a return. These came at the same time as spikes in the still pressure.
1731 Turn off MC heater and Still heater. Preparing for Hassan's suggested test for the still.
1748 Still temperature drops to lower than even before we began heating
1749 Still = 881 mK, MC = 558 mK
1751 Starting one-shot. Still temp increases to 955 mK, then begins dropping
1754 Still = 900 mK / 0.14 mbar. Target is 771 mK (100 lower than when we began). G2 = 30 mbar and rising as we pump mix to dump.
1755 Still = 850 mK / 0.11 mbar. G2 = 34 mbar.
1757 Still = 796 mK / 0.08 mbar. G2 = 37 mbar.
1757 Still = 781 mK, cease one-shot, resume circulation. Still temperature begins to increase.
1801 Still = 861 mK / 0.13 mbar.
1803 Still = 900 mK / 0.15 mbar, MC = 817 mK
1807 Still temperature decreases again and flattens out at 881 mK / 0.19 mbar – same temperature as before the one-shot. MC = 1.09 K
1810 begin recovery