| Time | G1 (mbar) | G2 (mbar) | P1 (ubar) | Still (K) | Additional Comment |
| 1215 | 124 | 111 | 220 | 0.964 | Commence test. Turn on Still Heater to 0.5 mW. |
| 1221 | 124 | 111 | 240 | 0.975 | Increase Still Heater to 1.0 mW. Mixing chamber temperature at 513 mK. |
| 1225 | 105 | 99 | 220 | 0.967 | Increase Still Heater to 2.0 mW. |
| 1230 | 102 | 96 | 240 | 0.980 | Mixing Chamber temperature at 477 mK and dropping. |
| 1236 | 102 | 96 | 260 | 0.985 | Increase Still Heater to 2.5 mW. |
| 1240 | 105 | 99 | 260 | 0.990 | Mixing Chamber temperature at 468 mK and dropping. |
| 1247 | 107 | 101 | 260 | 0.990 | Increase Still Heater to 3.0 mW. |
| 1249 | 109 | 103 | 260 | 0.997 | Mixing Chamber temperature at 458 mK and dropping. |
| 1304 | 113 | 107 | 260 | 0.997 | Increase Still Heater to 4.0 mW. |
| 1311 | 118 | 112 | 280 | 1.010 | Mixing Chamber temperature at 438 mK and dropping. |
| 1317 | 119 | 113 | 280 | 1.005 | Increase Still Heater to 5.0 mW. P1 initially rises to 300 ubar and then settles to 280 ubar. |
| 1324 | 123 | 111 | 280 | 1.018 | Mixing Chamber temperature at 421 mK and dropping. |
| 1330 | 126 | 118 | 280 | 1.017 | Increase Still Heater to 6.0 mW. P1 initially rises to 300 ubar. |
| 1341 | 134 | 126 | 300 | 1.028 | Mixing Chamber temperature at 406 mK and dropping. |
| 1347 | 139 | 131 | 300 | 1.024 | Increase Still Heater to 7.0 mW. P1 initially rises to 320 ubar. |
| 1353 | 145 | 136 | 320 | 1.035 | Mixing Chamber temperature at 394 mK and dropping. |
| 1400 | 130 | 127 | 320 | 1.037 | Increase Still Heater to 8.0 mW. P1 initially rises to 340 ubar and then settles to 320 ubar. |
| 1405 | 131 | 127 | 320 | 1.049 | Mixing Chamber temperature at 384 mK and dropping. |
| 1408 | 135 | 131 | 320 | 1.051 | Increase Still Heater to 9.0 mW. |
| 1420 | 152 | 145 | 340 | 1.087 | Mixing Chamber temperature at 373 mK and dropping. Still temperature is rising more quickly. |
| 1423 | 158 | 150 | 360 | 1.100 | Increase Still Heater to 10.0 mW. P1 initially rises to 380 ubar. |
| 1435 | 317 | 291 | 500 | 1.332 | Turn off Still Heater to avoid G1 and G2 rising higher. Mixing Chamber temperature at 333 mK and rising, though that is likely because the heater has been shut off. End test. |