| 1015 | Fill LN2 cold trap. |
| 1100 | Cease pumping on 1K pot, mixing chamber temperature begins rising from ~1.2K. Turn on mixing chamber heater and commence pumping to dump. |
| 1430 | Stop pumping on fridge lines by closing gate valve and V6. Current pressures: G1: 1.0mbar, G2: 649mbar, P1: 7ubar. |
| 1440 | Restart 1K pot. 1K Pot temperature ~1.7K. Commence condensation procedure. G2 reads 649mbar. Aiming for 300mbar condensing, so G2 should read 349mbar. |
| 1500 | Finish condensing. G1 and G2 at ~75mbar. Lower 1K pot temperature from 1.74K to 1.56K. Commence circulation procedure. |
| 1920 | Finish recording fridge temperature on mixing chamber thermometer, still thermometer, cold plate thermometer, and RuOx mixing chamber thermometer to document oscillatory behavior. Plots shown in Figures 1-4. NOTE: RuOx mixing chamber thermometer only records resistance, but an increase in the magnitude (i.e. a smaller resistance) corresponds to a lower temperature. |