| 1335 | At roughly 2300 last night, the mixing chamber thermometer reached a stable low point of 2.4 K. It had been cooling steadily from 32 K at roughly 2100 when the mixing chamber heater was unplugged. Thus it is likely that the mixing chamber heater was responsible for the high temperatures. Since that time, some mixture has been condensing, but the process is not completed. Currently mixing chamber temperature is 2.4 K, G1=90.6 mbar (cold trap 1 speedivalve closed), G2=315 mbar (V9 and V13A closed), P1=1000. Open coldtrap 1 speedivalve and V9 and V13A to continue circulation. Since 656 mbar was in G2 yesterday at the start of circulation, add 350 mbar to make G2 pressure 306 mbar. |
| 1545 | 306 mbar left in dump. Close V9. |
| 1550 | Fill LN2 coldtrap full. Start circulating procedure. G1=72.3 mbar, G2=40.7 mbar, mixing chamber at 2.23 K. |
| 1600 | V6 opened fully. G1=119 mbar, G2=104 mbar, P1=0.22 mbar, P2=26 mbar. Mixing chamber at 1.06 K and dropping. |
| 1602 | Mixing chamber at 875 mK and has starting to climb. |
| 1610 | Mixing chamber temperature at 1.16 K and still rising. P1=90 ubar, G1=71.5 mbar, G2=52 mbar. |
| 1621 | Plug in 1K Pot, reads 1.612 K. Unplug 1K Pot. Mixing chamber temperature still oscillating. |
| 1630 | Mixing chamber temperature drops from 1.1K to ~700 mK. |
| 1640 | Mixing chamber temperature reaches 1.129 K, G1= 90 mbar, G2=65 mbar. |
| 1648 | Mixing chamber temperature at 820 mK. |
| 1650 | Mixing chamber temperature maxes out at 900 mK. |
| 1651 | Mixing chamber temperature reaches a minimum of 850 mK. |
| 1658 | Mixing chamber temperature maxes out at 1.1 K. |
| 1659 | Mixing chamber temperature reaches a minimum of 980 mK. |
| 1702 | Mixing chamber temperature maxes out at 1.05 K. |
| 1704 | Mixing chamber temperature reaches a minimum of 697 mK. |
| 1706 | Mixing chamber temperature maxes out at 890 mK. |
| 1711 | Mixing chamber temperature reaches a maximum of 1.01 mK. These oscillations continue, but they will no longer be recorded unless there is a significant change. |
| 1754 | Mixing chamber temperature at 930 mK. Plug in 1K Pot thermometer/ mixing chamber heater. Stop circulation and start collecting mix in the dump. |