0910 Overnight external pressure sensor reached down to 5.4E-3 mbar (this might be lower, since this gauge tends to get stuck). Isolate turbopump, and see no immediate change in external pressure sensors.
0922 Still no change in external pressure. Turn on leak checker.
1005 Leak check pump room manifolds. Background leak rate of 8.8E-8 mbarL/s. Occasional spikes to 9.3 mbarL/s, but spikes were non-repeatable. No indication of leak on pumping manifolds.
1110 Raise dewar, pump out with scroll pump.
1200 Perform leak test on fridge lines by purging dewar with He gas. Initial leak rate on fridge lines was measured at 1.8E-7 mbarL/s. Adding He gas to the bath resulted in no obvious rise in leak rate; no leak detected.
1140 Start turbo pumping on condenser line and still line.
1245 Add He gas to still line to check throughout to condenser side. As soon as gas was added to still line, a change in pressure was registered on condenser line side, indicating a good (though not accurately measurable) throughput. Commence pumping on both sides of the fridge with the scroll pump and switch to turbo pump once the pressure is sufficiently low.
1510 Reconnect system to standard positioning. Pump system out with scroll pump. Switch to turbo pump.
1630 Perform throughput test with standard setup. Add He gas until G1 reads 570 mbar. Conduct throughput test by opening V1 (do not pump out G1 before hand; G1 drops to 46 mbar and then begins to rise. Throughput test gives a rate of 7 mbar/min.
1640 Commence pumping out of fridge lines with scroll pump. Once low enough, switch to turbo pump and leave overnight.