| 0925 | G1: 2mbar, P1: 17ubar, G2: 0mbar. External pressure sensor attached to He3 pump exhaust lines reads 4.0mbar. |
| 0948 | Fill LN2 cold trap full. |
| 1005 | Fill He gas through vent valve, V7, V11a, V12a till G1 reads 935mbar. P1: 16ubar. Open V1. |
| 1022 | G1: 839mbar, P1: 32ubar. |
| 1032 | G1: 849mbar, P1: 90ubar. Close LN2 cold trap 1 speedivalve because of rising G1 pressure. |
| 1037 | G1: 847mbar, P1: 150ubar. |
| 1041 | G1: 844mbar, P1: 190ubar. Add more He gas until G1 reads 999mbar. Open V1. |
| 1051 | G1: 876mbar, P1: 300ubar. |
| 1056 | G1: 871mbar, P1: 380ubar. Open V6. |
| 1108 | G1: 860mbar, P1: 500ubar. |
| 1118 | G1: 851mbar, P1: 400ubar. |
| 1126 | G1: 844mbar, P1: 380ubar. Open V2, V3 to aid in pumping out fridge lines. |
| 1145 | Isolate all valves on IGH. Open V7, V12a, cold trap 1 speedivalve, V13a, V3. G1: 732mbar. |
| 1155 | Test fails, pump out. G1: 608mbar, drops to a steady 47mbar. Does not rise, indicating no flow. |
| 1320 | G1: 637mbar drops and stabilizes at 45mbar indicating no flow. Pump out He gas. |
| 1520 | Add gas for “non-traditional” throughput test. G1: 561mbar, P1: 65ubar. Open V1, G1 drops to 43mbar. |
| 1535 | G1: 42mbar, P1: 75ubar. |
| 1607 | G1: 41mbar, P1: 90ubar. Pump out. Kill turbopump (only connection between pump and fridge lines passes through cold LN2 cold trap, which hopefully helps). Commence pumping out with rough pump. |
| 1745 | G1: 2mbar, P1: 60ubar. Add He gas to G1 through vent valve and LN2 cold trap 1 until G1 reaches 999mbar. Time to see if there's a new blockage in the fridge or if the behavior's the same. Open V1. |
| 1802 | G1: 910mbar, P1: 90ubar. Open V6 to see if pumping out through fridge is possible. |
| 1814 | G1: 900mbar, P1: 65ubar. |
| 1820 | G1: 894mbar, P1: 50ubar. Open V2 and V3 to aid in pumping. |
| 1847 | G1: 1.7mbar, P1: .40mbar. Isolate fridge lines. Lower dewar. |