1110 Come in to find that G2 has increased from 0 mbar to roughly 8.0 mbar. Checking the DAQ shows that this increase has been steadily occurring since about 7am this morning. G1=2.0 mbar and P1=18 ubar look normal, though that is to be expected since they are being pumped on by scroll pump. Current fridge temperature at 78 K. External pressure sensor reading He-3 pump exhaust is at 2.2E-5 mbar.
1130 Isolate He-3 pump from turbopump. External pressure sensor reads 1.2E-2 mbar. Turn off trubo pump. Lower dewar the rest of the way. Current fridge temperature at 79 K, and G2 now reads 11 mbar and is still rising.
1220 Close gate valve and condenser line speedivalve as well as all valves on IGH.
1240 Remove copper condenser line from the system and replace with He cold trap. Commence pumping on He-cold trap and IGH lines with scroll pump.
1550 G1: 1.8mbar, P1: 28ubar, G2: 8mbar, Mixing chamber temperature at 254K. External pressure sensor attached to He3 pump exhaust lines reads 2.2E-1mbar. Open gate and condenser line speedivalve to pump out fridge lines overnight to remove blockage.
P1 rises to 0.16 mbar.