| 1140 | Current Mixing Chamber temperature of 279K and G2 pressure of 502mbar. Commence pumping out IVC with turbopump. |
| 1145 | Fill LN2 cold trap full. |
| 1720 | Background helium leak rate 2.6E-9 mbar l/s. Commence spray test around 4 indium seals. |
| 1755 | Leak rate increased to 3.7E-9 mbar l/s very gradually. Doesn't look like the indium flanges are leaking. Move onto KF- and ISO-flanges above cryostat as these may have moved around a little and a hole may have opened up. |
| 1830 | Leak rate started to increase less gradually (but still gradually) when checking flange next to double-flange on top of wizard staff. Leak rate 1E-8 mbar l/s. Wait for leak rate to decrease. |
| 1950 | Leak rate has only just started to decrease having reached 1.1E-7 mbar l/s before falling again. |
| 2030 | Leak rate has been falling gradually, and has reached 5.5E-8 mbar l/s. Isolate leak checker and evacuate IVC with turbo pump. Let's try again tomorrow, starting at the flange next to the top of the wizard staff. |
| 2100 | Fill N2 cold trap full. |